Optimizing IVIM measurement on magnetic resonance linear accelerator using multi-b-value acquisition and physics-informed neural network.

Xinxiang Zhou, Na Huang, Zengyi Fang, Jie Tang, Tingting Zhang, Yulu Zhao, Lucia Clara Orlandini, Pei Wang,

Department of Radiation Oncology, Sichuan Clinical Research Center for Cancer, Sichuan Cancer Hospital & Institute, Sichuan Cancer Center, School of Medicine, University of Electronic Science and Technology of China, No. 55, Section 4, South Renmin Road, Chengdu, 610054, Sichuan, China; Department of Radiation Oncology, Precision Radiation in Oncology Key Laboratory of Sichuan Province, Sichuan Clinical Research Center for Cancer, Sichuan Cancer Hospital & Institute, Sichuan Cancer Center, University of Electronic Science and Technology of China, Chengdu, 610042, Sichuan, China. · Department of Radiation Oncology, Precision Radiation in Oncology Key Laboratory of Sichuan Province, Sichuan Clinical Research Center for Cancer, Sichuan Cancer Hospital & Institute, Sichuan Cancer Center, University of Electronic Science and Technology of China, Chengdu, 610042, Sichuan, China.

Computational

Key points

From the paper's conclusion
  • The combination of multi-b-value acquisition and PINN provided a robust and accurate framework for IVIM parameter estimation on the MR-Linac and showed potential for integration into MR-Linac quantitative imaging workflows.

Abstract

BACKGROUND: Accurate quantification of intravoxel incoherent motion (IVIM) parameters on the magnetic resonance linear accelerator (MR-Linac) is challenging due to low signal-to-noise ratio and time constraints. PURPOSE: This study aimed to evaluate the impact of b-value acquisition strategies and estimation algorithms on IVIM parameter measurement to guide reliable IVIM quantification on the MR-Linac. METHODS: A digital phantom with high number of signals averaged (high-NSA) and multi-b-value signals was constructed to evaluate estimation accuracy using root mean square error (RMSE). Ten healthy volunteers underwent four repeated diffusion-weighted imaging (DWI) scans, consisting of two high-NSA DWIs and two multi-b-value DWIs to assess repeatability via the within-subject coefficient of variation (wCV). IVIM parameters from high-NSA DWI were estimated using Segment fitting, while multi-b-value DWI data were processed using least squares (LS), Bayesian estimation, and physics-informed neural network (PINN). RESULTS: All IVIM parameters from multi-b-value DWI exhibited lower wCVs than those from high-NSA DWI, except for the LS-derived diffusion coefficient (D). Multi-b-value DWI combined with PINN achieved rapid and highly repeatable estimation. In the digital phantom, high-NSA DWI with Segment fitting yielded the highest RMSE for the pseudo-diffusion coefficient (D∗). LS and Bayesian methods resulted in higher D RMSEs and required longer fitting times. In contrast, multi-b-value DWI + PINN achieved the lowest RMSEs across all parameters with minimal computational cost. CONCLUSION: The combination of multi-b-value acquisition and PINN provided a robust and accurate framework for IVIM parameter estimation on the MR-Linac and showed potential for integration into MR-Linac quantitative imaging workflows.

Read the paper

Discussion