Diagnostic microbiology and infectious disease

90-Minute point-of-care PRATHAMASENSE genosensor for gene-specific diagnosis of infection and antimicrobial resistance in critically ill patients: a multicentre diagnostic accuracy study.

Kavita Khoiwal, Manoranjan Mahapatra, Balram Ji Omar, Pooja Goswami, Rishi Dhawan, Ankit Agarwal, Raghavendra Kumar Giri, Mukul Aggarwal,

Dept. of Obstetrics & Gynecology, All India Institute of Medical Sciences, Rishikesh, Uttarakhand, India. Electronic address: kavita.kh27@gmail.com. · Department of Hematology, All India Institute of Medical Sciences, New Delhi, India. Electronic address: mrmahapatra@hotmail.com.

Observational · n=33

Key points

From the paper's conclusion
  • PRATHAMASENSE Genosensor delivers gene-specific detection of bacteria and fungi, including AMR determinants, directly from blood, tracheal fluid, and urine within 90 minutes - without a culture or PCR step. This point-of-care platform meets the WHO REASSURED criteria. This study assessed diagnostic accuracy only.

Abstract

BACKGROUND: Infection remains a leading cause of morbidity and mortality when diagnosis and antimicrobial therapy are delayed. Standard culture testing requires 2-3 days for processing and reporting, forcing clinicians to initiate empirical broad-spectrum antibiotics - a major driver of antimicrobial resistance (AMR). We evaluated a novel point-of-care electrochemical DNA biosensor (PRATHAMASENSE Genosensor) against standard culture and molecular reference standards for the rapid detection of bacterial and fungal infections. METHODS: In this multicenter, cross-sectional diagnostic accuracy study conducted over 12 months (April 2024-January 2025) at AIIMS Rishikesh and AIIMS New Delhi, India, we tested 488 clinical specimens (232 blood, 10 tracheal fluid, 246 urine) from patients with acute febrile illness, sepsis, or febrile neutropenia. Genosensor results were compared with the standard culture (primary reference standard) and, in a representative subset (n = 33), with quantitative PCR (qPCR). Bayesian latent class analysis (BayesLCA, R) was used to estimate sensitivity and specificity against an unobserved latent infection status, given the known imperfect sensitivity of culture. We did not collect clinical outcomes or treatment data because the study focused on diagnostic accuracy. RESULTS: Against culture, the PRATHAMASENSE Genosensor showed 93% sensitivity (95% CI 82.3-99.6%) and 77.8% specificity (95% CI 52.2-98.5%) in blood/tracheal samples, and 91.3% sensitivity (95% CI78.2-99.6%) and 77.6% specificity (95% CI 54.6-99%) in urine. When benchmarked against a combined sample of blood and urine of culture vs. PRATHAMASESNE Genosensor, the 94.2% sensitivity (95% CI 85.4-99.7%) and 88.3% specificity (95% CI 68.7-99.6%) were obtained using BCLA analysis. However, as a molecular platform, the sample was tested in parallel both in PRATHAMASENSE Genosensor and Q-PCR (Biofire and RT-PCR), showing quite interesting results PRATHAMASENSE against qPCR (n = 33), sensitivity was 93.75% (95% CI 69.5-99.8), and specificity was 94.12% (95% CI 71.3-99.8%). PRATHAMASESNE Genosensor-positive/culture-negative cases showed clinical and biochemical profiles similar to those of culture-positive cases, consistent with a high false-negative rate for culture rather than a Genosensor specificity problem. CONCLUSION: PRATHAMASENSE Genosensor delivers gene-specific detection of bacteria and fungi, including AMR determinants, directly from blood, tracheal fluid, and urine within 90 minutes - without a culture or PCR step. This point-of-care platform meets the WHO REASSURED criteria. This study assessed diagnostic accuracy only.