Direct brainstem stimulation for intraoperative somatosensory evoked potential monitoring: A prospective study.
Kristina Lapteva, Andrey Gavrjushin, Aleksei Veselkov, Alena Chitadze, Anastasiia Kuznetsova, Shavkat Kadyrov
Laboratory of Clinical Neurophysiology, Burdenko Neurosurgery Center, 4th Tverskaya-Yamskaya Street 16, Moscow 125047, Russia. Electronic address: laptevakr@gmail.com. · Department of Glial Tumors, Burdenko Neurosurgery Center, 4th Tverskaya-Yamskaya Street 16, Moscow 125047, Russia. Electronic address: Agavrjushin@nsi.ru.
Key points
From the paper's conclusion- In this exploratory cohort, bsiSSEPs were feasible and consistently recordable, including in patients without recordable conventional pSSEPs. Although diagnostic performance estimates were favorable, differences in predictor definition and in-sample derivation of the bsiSSEP threshold preclude definitive conclusions regarding superiority over conventional pSSEPs. Further prospective validation using harmonized comparison criteria is required.
Abstract
OBJECTIVE: Intraoperative somatosensory evoked potential (SSEP) monitoring during brainstem and cervicomedullary surgery is limited by poor reproducibility and delayed feedback. We evaluated brainstem-stimulus-induced SSEPs (bsiSSEPs) and compared their diagnostic performance with conventional peripheral SSEPs (pSSEPs). METHODS: Twenty-three patients undergoing brainstem or cervicomedullary lesion resection were enrolled in this prospective single-center study. Both pSSEPs and bsiSSEPs were recorded intraoperatively. The primary outcome was postoperative worsening of deep sensation. Diagnostic performance was assessed using ROC analysis, McNemar's exact test, Cohen's κ, and DeLong's test. RESULTS: Conventional pSSEPs were recordable in 18/23 patients (78%), whereas bsiSSEPs were obtained in all patients with 15-30 averages. Using the exploratory ≥33% amplitude-reduction threshold, estimated sensitivity was 100% (95% CI 47.8-100.0%), specificity 94.4% (95% CI 72.7-99.9%), and AUC 0.978 (95% CI 0.927-1.000). In the paired subgroup, sensitivity and specificity were 100% and 92.9% for bsiSSEPs versus 75.0% and 50.0% for pSSEPs. Exploratory ROC analysis yielded AUCs of 0.964 and 0.625, respectively; however, predictor definitions differed. CONCLUSION: In this exploratory cohort, bsiSSEPs were feasible and consistently recordable, including in patients without recordable conventional pSSEPs. Although diagnostic performance estimates were favorable, differences in predictor definition and in-sample derivation of the bsiSSEP threshold preclude definitive conclusions regarding superiority over conventional pSSEPs. Further prospective validation using harmonized comparison criteria is required. SIGNIFICANCE: Direct brainstem or upper cervical spinal cord stimulation addresses a major limitation of conventional sensory monitoring by providing rapid scalp-recorded evoked responses without relying on peripheral nerve or the sublesional portion of the spinal somatosensory pathway.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Intraoperative Neurophysiological Monitoring During Adult Brainstem Glioma Resection: Association Between Electrophysiological Changes and Early Neurological Outcome.Current oncology (Toronto, Ont.) · 2026
- Sequential intraoperative protocol to improve bulbocavernosus reflex monitorability during pediatric spinal surgery.Child's nervous system : ChNS : official journal of the International Society for Pediatric Neurosurgery · 2026
- Continuous facial corticobulbar motor evoked potential monitoring during vestibular schwannoma surgery: predictive value for postoperative facial nerve function.Frontiers in surgery · 2026
- 1038 Diagnostic Accuracy of Intraoperative Facial Motor Evoked Potentials for Predicting Postoperative Facial Dysfunction After Vestibular Schwannoma Surgery: A Systematic Review and Meta-AnalysisNeuro-Oncology Advances · 2026
- Standards for Reporting of Diagnostic Accuracy involving Intraoperative Neurophysiological MonitoringmedRxiv · 2026
- Detection and Early Detection of Pathological Motor Evoked Potentials: A Novel Approach for Intraoperative Neuromonitoring Assistance.Journal of healthcare informatics research · 2026