Journal of chromatography. A

Absolute, fast, accurate and easy-to-use method for determining the injected samples volumes in capillary zone electrophoresis using an injection tracer.

Farid Oukacine, Luc Choisnard, Lyn Boukai, Eric Peyrin

DPM, Grenoble Alpes University, CNRS UMR 5063, F-38041, Grenoble, France. Electronic address: farid.oukacine@univ-grenoble-alpes.fr. · DPM, Grenoble Alpes University, CNRS UMR 5063, F-38041, Grenoble, France.

Key points

From the paper's conclusion
  • The method was further validated by performing sequential injections of multiple samples, demonstrating high precision across all injections.

Abstract

In this study, a simple, rapid, and reliable methodology was developed for measuring the injected sample volume in capillary electrophoresis. The method is compatible with hydrodynamic injection and does not require a calibration curve, nor knowledge of sample viscosity, injection pressure, or time. It is applicable under both significant and weak electroosmotic flow and is compatible with single or sequential injections. The principle of the method is straightforward: a band of neutral molecules is injected before the sample, which has been previously spiked with the same molecule. Classical mechanics allow the absolute calculation of the injected band volume and the area of this band is simply compared with that of the neutral molecule peak to determine the injected volume of the unknown sample. Experimental validation demonstrated excellent agreement with volumes calculated from Poiseuille's law. The method was further validated by performing sequential injections of multiple samples, demonstrating high precision across all injections.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers