Rapid DNA fragment sizing by short-path microchip electrophoresis with CMOS fluorescence imaging.
Zhenqing Li, Xiaoyu Zhu, Bo Yang, Jing Yang, Dawei Zhang, Yoshinori Yamaguchi, Shengqiong Deng
Engineering Research Center of Optical Instrument and System, Ministry of Education, Shanghai Key Lab of Modern Optical System, University of Shanghai for Science and Technology, No 516 JunGong Road, Shanghai, 200093, China. · Faculty of Engineering, Anhui Sanlian University, Hefei, 230601, China. Electronic address: yangjing@mail.slu.edu.cn.
Key points
From the paper's conclusion- Under these conditions, the system exhibited a clear concentration-dependent fluorescence response over 0.04-2.0 ng μL⁻¹, with an estimated detection limit of 0.04 ng μL⁻¹ (S/N ≈ 3) for a 100 bp fragment, and PCR amplicons from Porphyromonas gingivalis and pEGFP-N1 were detected within approximately 30 s.
Abstract
Rapid post-PCR DNA fragment analysis requires short separation times while maintaining sufficient sizing capability. Short-path microchip electrophoresis (MCE) offers a direct route to accelerate electrophoretic analysis, but reducing the separation distance compresses the migration window and places greater demands on fluorescence detection and signal extraction. In this work, a short-path MCE platform coupled with CMOS fluorescence imaging was developed for rapid DNA fragment sizing. Sequential fluorescence images acquired at a fixed detection region were converted into intensity-time electropherograms, enabling migration analysis within a compact optical configuration. The effects of sieving matrix, fluorescent dye, effective separation length, and electric-field strength were systematically investigated, leading to the selection of 0.5% HEC, SYBR Green I, a 10 mm effective separation length, and a separation voltage of 535 V. Under these conditions, the system exhibited a clear concentration-dependent fluorescence response over 0.04-2.0 ng μL⁻¹, with an estimated detection limit of 0.04 ng μL⁻¹ (S/N ≈ 3) for a 100 bp fragment, and PCR amplicons from Porphyromonas gingivalis and pEGFP-N1 were detected within approximately 30 s. Using 50 and 500 bp DNA marker fragments as reference points, the apparent sizes of the two PCR products were estimated to be 79.4 ± 4.1 and 157.8 ± 5.3 bp, respectively. The proposed approach provides a rapid and compact platform for post-PCR fragment analysis and size estimation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Microfluidic impedance cytometry chip combining composite cell pre-focusing microchannel and 3D symmetric microelectrodes for cell classificationBiomicrofluidics · 2026
- Parallel fluorescence interrogation of multichannel chips for rapid point-of-care testing.Analytical and bioanalytical chemistry · 2026
- CircDNA-PAINT: high-resolution binary-barcode multiplexing with circular DNA docking strandsbioRxiv · 2026
- Versatile Microfluidics Platform for Enhanced Multitarget Super-Resolution MicroscopyACS Nano · 2026
- A compact centrifugal multi-volume digital PCR disk (MV-dPCR disk) with integrated pneumatic stabilization.Lab on a chip · 2026
- Microfluidics-Enhanced 3D DNA-Based Points Accumulation for Imaging in Nanoscale Topography.ACS applied materials & interfaces · 2026