Optimization and application of a sensitive UPLC-MS/MS method for the quantitative determination of antibiotics in PM2.5.
Xinyue Bei, Juan Liu, Qiao Yao, Xiao Ma, Shuyi Zhang, Yida Chen, Chao Wang, Wen Gu,
Chinese Center for Disease Control and Prevention Key Laboratory of Environment and Population Health, National Institute of Environmental Health, Chinese Center for Disease Control and Prevention & Chinese Academy of Preventive Medicine, Beijing, 100021, China. · Chinese Center for Disease Control and Prevention Key Laboratory of Environment and Population Health, National Institute of Environmental Health, Chinese Center for Disease Control and Prevention & Chinese Academy of Preventive Medicine, Beijing, 100021, China; National Key Laboratory of Intelligent Tracking and Forecasting for Infectious Diseases, National Institute of Environmental Health, Chinese Center for Disease Control and Prevention & Chinese Academy of Preventive Medicine, Beijing, 100021, China. Electronic address: juanliu674511@126.com.
Key points
From the paper's conclusion- Compared with previously reported methods that generally cover fewer antibiotic classes or focus on conventional environmental matrices, the proposed method enables sensitive and simultaneous determination of a broad range of antibiotics in the analytically challenging PM2.5 matrix.
Abstract
The increasing occurrence of antibiotics in the environment has raised concerns about their atmospheric transport, human exposure, and potential contribution to antibiotic resistance dissemination. In this study, a sensitive ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method was developed and validated for the simultaneous determination of 74 antibiotics from multiple classes in fine particulate matter (PM2.5). Key chromatographic and sample-preparation parameters were systematically optimized. Optimal performance was achieved using a T3 chromatographic column, mobile phases containing 0.1% (v/v) formic acid, Oasis HLB cartridges for solid-phase extraction, 0.1 mol/L disodium ethylenediaminetetraacetate, and methanol containing 1% (v/v) formic acid as the elution solvent. These conditions improved extraction efficiency, chromatographic performance, and quantitative sensitivity while reducing matrix-related interference. The validated method yielded accuracy values of 69.7%-131.7% and relative standard deviations of 1.0%-18.1% at three fortification levels, with estimated method quantification limits as low as 0.37 pg/m3. The method was applied to 29 ambient PM2.5 samples collected over 24-h periods in Beijing. Twenty-two antibiotics were detected above their respective method detection limits, with concentrations reaching 121.38 pg/m3. Seven compounds exhibited detection frequencies above 50%, and roxithromycin was detected in all samples. Oxytetracycline showed the highest mean concentration among the quantifiable results, at 14.48 pg/m3. Compared with previously reported methods that generally cover fewer antibiotic classes or focus on conventional environmental matrices, the proposed method enables sensitive and simultaneous determination of a broad range of antibiotics in the analytically challenging PM2.5 matrix. It therefore provides a comprehensive workflow for monitoring airborne antibiotics and supporting subsequent exposure and environmental-occurrence studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Use of direct injection UHPLC-MS/MS for antibiotic population use assessment by wastewater analysis - application to laboratory and monitoring studies.Analytical methods : advancing methods and applications · 2026
- Optimization and validation of a two-step extraction dispersive liquid–liquid microextraction (UA-DLLME) method for the determination of 17 antibiotics in influent wastewaterJournal of Chromatography A · 2026
- Robotic-assisted magnetic dispersive solid-phase extraction coupled with UHPLC-MS/MS for antibiotic analysis in aqueous matrices.Journal of chromatography. A · 2026
- Development and validation of a multi-class solid-phase extraction and liquid chromatography tandem mass spectrometry method for emerging contaminants in waterJournal of Chromatography A · 2025
- Sustainable UHPLC-MS/MS assay for ultra-sensitive simultaneous quantification of amoxicillin and flucloxacillin with application to a bioequivalence study.Analytical and bioanalytical chemistry · 2026
- Robust ultra-trace quantification of multi-class antibiotics in environmental waters using automated 96-well SPE and LC-HRMS.Journal of chromatography. A · 2026