A mini SPE purification method for the simultaneous determination of 15 mycotoxins in Ginkgo biloba leaves and their extracts by UHPLC-MS/MS.
Congmin Liu, Xinjia Tong, Qing Miao, Li Li, Jin Ye, Zhihong Xuan, Baoxia Ni, Yu Wu,
College of Chemistry and Pharmaceutical Sciences, Qingdao Agricultural University, Qingdao, 266109, China. · College of Chemistry and Pharmaceutical Sciences, Qingdao Agricultural University, Qingdao, 266109, China; Academy of National Food and Strategic Reserves Administration, No.11 Baiwanzhuang Street, Xicheng District, Beijing, 100037, China.
Key points
From the paper's conclusion- An analytical method combining isotope-dilution ultra-high-performance liquid chromatography coupled with mass spectrometry (ID-UHPLC-MS/MS) with mini solid phase extraction (mini SPE) was established and validated for the simultaneous determination of 15 mycotoxins in Ginkgo biloba leaves and their extracts.
Abstract
An analytical method combining isotope-dilution ultra-high-performance liquid chromatography coupled with mass spectrometry (ID-UHPLC-MS/MS) with mini solid phase extraction (mini SPE) was established and validated for the simultaneous determination of 15 mycotoxins in Ginkgo biloba leaves and their extracts. The mini SPE column, containing only one-quarter of the sorbent used in a conventional SPE cartridge, maintained satisfactory purification performance while reducing the excessive purification scale of conventional SPE relative to the final sample volume prepared for UHPLC-MS/MS analysis, thereby decreasing sorbent and solvent consumption. Optimized conditions yielded excellent linearity (R² >0.99), acceptable recoveries (73.67%-129.05%) and precision (RSDs <20.25%). Its limits of detection ranged from 0.34 μg/kg to 272.72 μg/kg. Actual samples showed low contamination of DON, DON-3 G, 3-AcDON, 15-AcDON and FB₁. In vitro assays confirmed Ginkgo biloba leaves ethanol extracts inhibited three key toxigenic fungi, possibly reducing mycotoxin levels. This method not only provided a cost-effective and reliable monitoring tool for mycotoxin detection in Ginkgo biloba leaves and extracts, but also offered valuable insights and references for the miniaturization development of other SPE purification columns.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multi-Metric Assessment of Liquid Chromatography-Mass Spectrometry Methods for Mycotoxin Determination in Human Urine: Advancing Sustainable, Practical, and Innovative Analytical Practices.Pharmaceuticals (Basel, Switzerland) · 2026
- A universally automated and ultra-efficient pretreatment device for rapid assay of mycotoxins in grain and oil.Food chemistry · 2026
- Mycotoxin occurrence and risk assessment in plant-based meat, cheese, and fish alternatives based on an adapted UHPLC-MS/MS multi-methodMycotoxin Research · 2026
- Online solid phase extraction-LC-MS/MS with two step peak focusing for sensitive multi-analyte analysis of mycotoxins in urineMicrochemical Journal · 2025
- Fully Automated Purification of Small Molecules Using Online MicroSPE-UPLC-SPE.Analytical chemistry · 2026
- Development of a Sensitive UPLC-MS/MS Method for the Simultaneous Quantification of Mycotoxins in Wheat Products and Human UrineToxins · 2026