High-Capacity transcranial Direct Current Stimulation (HC-tDCS).
Kyle Donnery, Mohamad FallahRad, Santiago Osorio, Mojtaba Belali Koochesfahani, Niranjan Khadka, Matthew Saw, Jinglin Mo, Benjamin Babaev,
Department of Biomedical Engineering, The City College of New York, CUNY, New York, NY. Electronic address: kdonnery@ccny.cuny.edu. · Department of Biomedical Engineering, The City College of New York, CUNY, New York, NY.
Key points
From the paper's conclusion- HC-tDCS was well tolerated by participants and expands the accessible tDCS dose range, enabling stimulation up to at least 6 mA and substantially higher electric fields throughout cortical and deeper brain regions than conventional 2 mA tDCS.
Abstract
BACKGROUND: Enhancing tDCS technology can support the delivery of higher current intensities, enabling broader dose-response studies in human clinical trials. METHODS: High-Capacity tDCS (HC-tDCS) integrates novel electrodes and adaptive current/voltage controlled electronics. Multi-layer HC electrodes include polarity-specific redox layers and engineered hydrogel interfaces, configured for a bifrontotemporal montage. The stimulator design includes adaptive ramps with hybrid voltage-current control and low compliance voltage. Scanning electron microscopy (SEM) and electrical impedance spectroscopy (EIS) characterized electrode properties. Tolerability of HC-tDCS was tested at target currents of 1-6 mA (in 1 mA increments) for 30 min on 10 healthy subjects, and compared with conventional tDCS using 2 mA F3-F4 sponge-electrodes. Tolerability was assessed according to the 100 mm visual analogue scale for pain (VAS-P100), skin erythema assessment, thermal imaging, and adverse event questionnaires. MRI-derived computational models predicted brain electric fields. RESULTS: The electrode design incorporating roughened redox layers with polarity-specific electrochemical capacity supports high-charge direct current stimulation. In all subjects, HC-tDCS was well tolerated at all tested doses (1-6 mA) with minor transient adverse events. Mean VAS-P100 (± SD) was 17.17 ± 11.12 for 6 mA HC-tDCS and 18.30 ± 9.49 for 2 mA sponge-electrode tDCS. HC-tDCS operated at lower voltage than sponge-tDCS. Across MRI-derived head models, 6 mA HC-tDCS produced 2.25-3.26 V/m electric fields, ∼4x higher than conventional 2 mA tDCS electric fields of 0.49-0.85 V/m. 6 mA HC-tDCS produced >1 V/m across all evaluated regions, including dorsolateral prefrontal cortex, anterior cingulate cortex, insula, and amygdala, whereas no evaluated region reached 0.5 V/m with conventional 2 mA tDCS. CONCLUSIONS: HC-tDCS was well tolerated by participants and expands the accessible tDCS dose range, enabling stimulation up to at least 6 mA and substantially higher electric fields throughout cortical and deeper brain regions than conventional 2 mA tDCS.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accelerated high-capacity 6 mA transcranial direct current stimulation for major depression.medRxiv : the preprint server for health sciences · 2026
- Computational analysis of a high-density electrode to scale the electric field geometry and intensity for neurostimulation.Biomedical engineering online · 2026
- From Protocol to Practice: Hidden Sources of Variability in Transcranial Direct Current Stimulation.Journal of visualized experiments : JoVE · 2026
- Comparative electrothermal analysis of single and small-electrode grid configurations for HD-inspired peripheral transcutaneous neurostimulation.Frontiers in bioengineering and biotechnology · 2026
- Neural mechanisms of tDCS: Insights from an in-vivo rodent model across a range of electric field strengthsBrain stimulation · 2026
- Assessing the effect of tDCS on inhibitory capacity: A pre-registered comparison of high definition and conventional montages on the stop signal taskCortex · 2026