Journal of plastic, reconstructive & aesthetic surgery : JPRAS

Bipolar intraoperative nerve stimulation shows distinct response patterns compared with monopolar stimulation after median nerve stretch injury in rats.

Christoph A Schroen, Michael R Hausman, Paul J Cagle

Leni & Peter W. May Department of Orthopaedic Surgery, Icahn School of Medicine at Mount Sinai, New York, NY, United States; Medical Faculty Heidelberg, Heidelberg University, Heidelberg, Germany. Electronic address: christoph.schroen@googlemail.com. · Leni & Peter W. May Department of Orthopaedic Surgery, Icahn School of Medicine at Mount Sinai, New York, NY, United States.

Preclinical

Key points

From the paper's conclusion
  • Bipolar stimulation detected acute nerve injury in a rodent model but did not discriminate injury severity or recovery potential. Monopolar stimulation retained responses in many recoverable injuries, suggesting that the two devices are not interchangeable for intraoperative prognostic assessment.

Abstract

BACKGROUND: Intraoperative electrical nerve stimulation may help assess the severity and recovery potential of incontinuity peripheral nerve injuries. Although monopolar stimulation has shown prognostic value after nerve stretch injury, the utility of bipolar stimulation remains unclear. This study compared handheld monopolar and bipolar nerve stimulators for acute assessment of recoverable and nonrecoverable stretch injuries in rats. METHODS: Twenty-two male Sprague-Dawley rats underwent bilateral median nerve exposure. Nerves were assigned to sham control, epineuroclasis, or endoneuroclasis injury. Epineuroclasis represents the first mechanical failure event during stretching and is associated with recovery, whereas endoneuroclasis represents a more severe injury with poor functional recovery. Nerves were tested intraoperatively with commercially available monopolar and bipolar stimulators before and after injury. Postinjury responses were analyzed as binary outcomes, based on visible digit flexion. RESULTS: Bipolar stimulation distinguished injured from uninjured nerves, with responses in 12 of 12 sham-control nerves compared with 4 of 32 injured nerves. However, bipolar stimulation did not distinguish recoverable epineuroclasis from nonrecoverable endoneuroclasis injuries. In contrast, discordant responses after epineuroclasis occurred exclusively in one direction: 12 nerves responded to monopolar but not bipolar stimulation, whereas no nerve responded to bipolar stimulation alone. CONCLUSION: Bipolar stimulation detected acute nerve injury in a rodent model but did not discriminate injury severity or recovery potential. Monopolar stimulation retained responses in many recoverable injuries, suggesting that the two devices are not interchangeable for intraoperative prognostic assessment.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Bipolar intraoperative nerve stimulation shows distinct response patterns compared with monopolar stimulation after median nerve stretch injury in rats — Science Skim