Biosensors & bioelectronics

Fluorescent sandwich immunoassay microfluidic platform integrating a CGQDs-functionalized interface and W/V microstructures for multiplex quantitative detection of food allergens.

Xin Zhao, Sijia Wang, Siddanakoppalu N Pramod, Yonghong Li, Jian Huang, Lichun Liu, Hong Lin, Zhenxing Li

College of Food Science and Engineering, Ocean University of China, Sansha Road 1299, Qingdao, 266404, China. · Laboratory of Immunomodulation and Inflammation Biology, Department of Studies and Research in Biochemistry, Sahyadri Science College, Kuvempu University, Shimoga 577203, Karnataka, India.

Key points

From the paper's conclusion
  • Under optimized conditions, the platform completed the in-channel immunoassay process within 40 min after antigen introduction, with limits of detection (LODs) of 2.74-14.32 ng/mL and limits of quantification (LOQs) of 6.55-44.45 ng/mL.

Abstract

Accurate identification of allergens is essential for food safety control and risk management, yet simultaneous multicomponent analysis requires efficient analytical strategies. In this study, a fluorescent sandwich immunoassay microfluidic platform (FSIMP) integrating an interface functionalized with carboxylated graphene quantum dots (CGQDs) and channels incorporating W- and V-shaped (W/V) microstructures was constructed for simultaneous quantitative detection of sarcoplasmic calcium-binding protein (SCP), ovalbumin (OVA), tropomyosin (TM), walnut protein (WP), and peanut protein (PP). Finite element simulations characterized flow modulation by W/V microstructures, while experiments showed signal-to-noise (S/N) ratios 1.42-1.77 times those of planar channels. Positive-to-negative (P/N) values were 2.9%-84.4% higher on CGQDs substrates than on amino-modified slides. Under optimized conditions, the platform completed the in-channel immunoassay process within 40 min after antigen introduction, with limits of detection (LODs) of 2.74-14.32 ng/mL and limits of quantification (LOQs) of 6.55-44.45 ng/mL. Intra- and inter-chip coefficients of variation were 5.56%-11.82% and 2.77%-8.06%, respectively, with spike recoveries of 84.45%-142.40%. Mean fluorescence differed by ≤ 8.09% between days and between chip batches. After 3 days at 37 °C, P/N ratios were 61.20%-99.16% of those in controls stored at room temperature for the same period. Nine commercial foods were analyzed. For OVA and TM in six foods, three of seven comparisons with commercial ELISA showed relative differences of 2.28%-7.10%; in the other four, allergen concentrations measured by FSIMP were 2.75-17.89 times the corresponding ELISA values. This study provides a microfluidic approach for simultaneous quantification of multiple allergens in foods.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.