Biosensors & bioelectronics

Cascade catalytic self-supply of •OH coreactants enabled by GOx/Mn/Co-MOF hybrids for enhanced electrochemiluminescence sensing of miRNA-222.

Yan Zhang, Yongguang Gao, Ying Zhang, Xiangmin Miao, Yuehua Guo

Research Center of Clinical Medicine, Affiliated Hospital of Nantong University, Medical School of Nantong University, Nantong, 226001, China. · Department of Radiology, Xuzhou Central Hospital, Southeast University Affiliated Xuzhou Central Hospital, Xuzhou, 221000, China.

Key points

From the paper's conclusion
  • Here, we developed a strategy based on the cascade catalytic self-supply of a •OH coreactant to enhance ECL sensing for miRNA-222, achieved via a cascaded self-feedback circuit and catalytic hairpin assembly (CHA).

Abstract

In conventional electrochemiluminescence (ECL) sensing systems, the use of H2O2 as a coreactant often suffers from poor stability, susceptibility to decomposition, and weak reproducibility. Accordingly, constructing cascade autocatalytic reactions to in-situ generate hydroxyl radicals (•OH) is of great significance. Here, we developed a strategy based on the cascade catalytic self-supply of a •OH coreactant to enhance ECL sensing for miRNA-222, achieved via a cascaded self-feedback circuit and catalytic hairpin assembly (CHA). Specifically, a ternary hybrid nanozyme comprising highly dispersed manganese species and glucose oxidase (GOx) anchored on a hollow-structured cobalt metal-organic framework (GOx/Mn/Co-MOF) was constructed to catalyze glucose for the generation of •OH. The resulting •OH reacted with luminol-functionalized gold nanoparticles (L-Au NPs) to yield amplified ECL emission. Impressively, the integration of the cascaded self-feedback circuit and CHA enabled an efficient DNA cascade reaction, realizing remarkable tandem signal amplification and further enhancing the detection sensitivity for miRNA-222. As a result, the constructed biosensor achieved the detection of miRNA-222 over a wide linear range from 100 aM to 100 pM, with a detection limit down to 30 aM. More importantly, this method realized the accurate analysis of miRNA-222 in serum samples, offering a reliable analytical platform for early disease screening and clinical diagnosis.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Cascade catalytic self-supply of •OH coreactants enabled by GOx/Mn/Co-MOF hybrids for enhanced electrochemiluminescence sensing of miRNA-222 — Science Skim