Brain & spine

Usage and pitfalls of intracranial/invasive multimodal neuromonitoring during MR imaging of neurocritical care patients: a questionnaire study.

Daniel Pinggera, Philipp Geiger, Victoria Schön, Marlies Bauer, Christian Preuss-Hernandez, Sami Barrit, Mejdeddine Al Barajraji, Bart Depreitere,

Department of Neurosurgery, Medical University of Innsbruck, Innsbruck, Austria. · Computational Precision Health, University of California, Berkeley, CA, 94720-2284, United States.

Key points

From the paper's conclusion
  • Ninety clinicians from 29 countries participated, predominantly neurosurgeons (85.6%) practicing in Europe (71/84, 84.5%). Frontal placement was reported by 97.1% (66/68) and ipsilateral placement by 67.2% (43/64). Probes were inserted in the operating room by 51.5% and at the bedside by 48.5%; 61.8% used bolt systems (n = 68). Of 87 respondents, 51.7% kept all probes in situ during MRI, 29.9% removed all and 18.4% removed selected probes. Respondents removing probes (n = 41) most often cited absent MRI safety testing (46.3%) and thermal injury (41.5%); of those leaving probes in situ (n = 57), 66.7% reported no complications. MRI was performed on demand by 54.0% and within seven days by 37.9% (n = 87). No respondent characteristic was significantly associated with these practices.

Abstract

INTRODUCTION: Invasive intracranial neuromonitoring and early magnetic resonance imaging (MRI) inform treatment and prognosis in traumatic brain injury (TBI), but probes in situ raise MRI safety concerns. RESEARCH QUESTION: How do neurocritical care clinicians handle MRI in patients with intracranial neuromonitoring probes? MATERIALS AND METHODS: A cross-sectional open online survey was distributed via European and international neurosurgical and neurocritical care networks. It covered demographics, probe placement and MRI practices. Data were analyzed descriptively with 95% confidence intervals and exploratory Fisher's exact tests. RESULTS: Ninety clinicians from 29 countries participated, predominantly neurosurgeons (85.6%) practicing in Europe (71/84, 84.5%). Frontal placement was reported by 97.1% (66/68) and ipsilateral placement by 67.2% (43/64). Probes were inserted in the operating room by 51.5% and at the bedside by 48.5%; 61.8% used bolt systems (n = 68). Of 87 respondents, 51.7% kept all probes in situ during MRI, 29.9% removed all and 18.4% removed selected probes. Respondents removing probes (n = 41) most often cited absent MRI safety testing (46.3%) and thermal injury (41.5%); of those leaving probes in situ (n = 57), 66.7% reported no complications. MRI was performed on demand by 54.0% and within seven days by 37.9% (n = 87). No respondent characteristic was significantly associated with these practices. DISCUSSION AND CONCLUSION: Respondents broadly agreed on frontal placement, whereas lateralization, insertion technique and MRI-related probe handling varied considerably. These self-reported data from a predominantly European, neurosurgical sample cannot establish MRI safety with probes in situ. Device-specific MRI safety testing, prospective multicenter registries and multidisciplinary consensus are needed to standardize practice.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers