Safety And Efficacy of Microfibrillar Collagen Hemostat in Carotid Artery Heparinized Bleeding Injury Ovine model compared to Gelatin-thrombin Hemostatic Matrix.
Amie Rossi, Darcy Gagne, Kasia Bradbury, John E Mayer
Becton, Dickinson and Company, Warwick, Rhode Island, USA. · Becton, Dickinson and Company, Warwick, Rhode Island, USA. Electronic address: Darcy.Gagne@bd.com.
Key points
From the paper's conclusion- In this ovine model, Flowable Collagen achieved effective hemostasis with no device-related adverse effects identified, and hemostatic outcomes similar to Gelatin Matrix. Lower overall inflammation scores observed for Flowable Collagen at Day 120 require confirmation in adequately powered studies.
Abstract
INTRODUCTION: Rapid and effective hemostasis is critical in cardiovascular surgeries. This study assesses the hemostatic efficacy and safety of Flowable Microfibrillar Collagen Hemostat compared to Gelatin-thrombin Matrix when applied over an ePTFE (expanded polytetrafluoroethylene cardiovascular patch) carotid artery patch in heparinized vascular bleeding injury ovine model. METHODS: Twelve heparinized Polypay sheep underwent unilateral carotid artery incision and repair with ePTFE carotid artery patch. Suture lines were treated with Flowable Collagen (n=6) or Gelatin Matrix (n=5). Bleed grade, tamponade time, and hemostasis at 10 and 30 minutes were recorded. Angiographic assessments and tissue evaluations were performed at Day 0 (baseline and post-treatment), at Day 60 and Day 120. RESULTS: Mean bleed grade was 1.8 for both groups prior to treatment. Mean total tamponade time was 3.7 minutes for Flowable Collagen and 4.0 minutes for Gelatin Matrix. At 10 minutes, hemostasis was achieved in 83% and 80% of treated grafts, respectively. All sites were hemostatic at 30 minutes. No significant differences were observed. Gelatin Matrix-treated arteries demonstrated greater reduction in the mean-distal diameter post-treatment (7.82 mm to 6.93 mm) than Flowable Collagen-treated arteries (7.88 mm to 27 mm). On Day 120, overall inflammation was significantly lower in Flowable Collagen treated arteries when compared to Gelatin Matrix. CONCLUSION: In this ovine model, Flowable Collagen achieved effective hemostasis with no device-related adverse effects identified, and hemostatic outcomes similar to Gelatin Matrix. Lower overall inflammation scores observed for Flowable Collagen at Day 120 require confirmation in adequately powered studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Injectable biochar-reinforced alginate hydrogels for hemostasis under wet and deformable conditions.Colloids and surfaces. B, Biointerfaces · 2026
- Hierarchically sized chitosan hemostatic particles with electrostatic-hydrophobic synergy for rapid and effective control of arterial hemorrhage.Colloids and surfaces. B, Biointerfaces · 2026
- Sutureless repair of inferior vena cava injury using an elastomeric sealant in a mouse model with histopathological analysis of vascular healing.JVS-vascular science · 2026
- Analysis of Hyaluronic Acid (HA), Calcium Hydroxylapatite (CaHA), HA + CaHA Embolization in Rabbit Ears and Thrombolysis Recovery.Aesthetic plastic surgery · 2026
- Application of gelatin sponge (Gelfoam®) as a hemostatic agent in inflatable penile prosthesis implantationInternational Journal of Impotence Research · 2025
- In vitro characterization and in vivo technical feasibility of biphasic GelMA/HAMA/hydroxyapatite cell-laden hydrogels for osteochondral regeneration: a pilot study in an ovine modelFrontiers in Bioengineering and Biotechnology · 2026