Early and midterm outcomes of the Gore Excluder Iliac Branch Endoprosthesis within and outside the instructions for use for aortoiliac and iliac artery aneurysms.
Masayuki Hara, Kentaro Kasa, Takao Ohki, Kota Shukuzawa, Soichiro Fukushima, Hirotsugu Ozawa, Makiko Omori, Miyo Shirouzu,
Division of Vascular Surgery, Department of Surgery, The Jikei University School of Medicine, 3-25-8, Nishi-shimbashi, Minato-ku, Tokyo, 105-8461, Japan. Electronic address: haharappy@hotmail.com. · Division of Vascular Surgery, Department of Surgery, The Jikei University School of Medicine, 3-25-8, Nishi-shimbashi, Minato-ku, Tokyo, 105-8461, Japan. Electronic address: kasa@jikei.ac.jp.
Key points
From the paper's conclusion- IBE use outside the IFU was feasible in carefully selected patients. No significant differences in midterm clinical outcomes were detected between the within-IFU and outside-IFU groups.
Abstract
OBJECTIVE: To evaluate perioperative and midterm outcomes of endovascular aneurysm repair using the Gore Excluder Iliac Branch Endoprosthesis (IBE) for aortoiliac and iliac artery aneurysms treated within and outside the instructions for use (IFU). METHODS: This single-center retrospective cohort study included patients who underwent IBE implantation for non-ruptured aortoiliac or iliac artery aneurysms between April 2017 and December 2022. Outcomes were analyzed at the patient, procedure, or IBE-treated-side level, as appropriate. The primary endpoints were technical success, IBE-related complications, and reinterventions. Secondary endpoints included internal iliac artery (IIA) branch patency, aneurysm sac status, and freedom from reintervention or aneurysm-related death. RESULTS: A total of 56 unique patients underwent 58 procedures involving 73 IBE-treated sides; 24 sides were within-IFU and 49 were outside-IFU. Technical success was achieved in 100% of within-IFU procedures and 95.0% of outside-IFU procedures. At 36 months, freedom from reintervention was 79.1% and 94.0% in the within-IFU and outside-IFU groups, respectively (P = .24), and IIA branch patency was 88.8% and 85.5%, respectively. No significant between-group difference in IIA branch patency was detected after accounting for within-patient clustering (hazard ratio, 1.53; 95% confidence interval, 0.28-8.35; P = .62). Seven IIA branch occlusions occurred during follow-up, none of which was associated with buttock claudication. In exploratory univariable analysis, type II endoleak was associated with CIA aneurysm sac growth (odds ratio, 7.86; 95% confidence interval, 1.08-57.0; P = .04). CONCLUSIONS: IBE use outside the IFU was feasible in carefully selected patients. No significant differences in midterm clinical outcomes were detected between the within-IFU and outside-IFU groups.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- International Multicentre Study on mid- and long-term Outcome of E-liac Iliac Branched Device.Cardiovascular and interventional radiology · 2026
- Early Postoperative Simultaneous Occlusion of the Internal and External Iliac Artery Limbs after GORE Excluder Iliac Branch Endoprosthesis Implantation.Annals of vascular diseases · 2026
- Early Outcomes of Physician-Modified Endurant Limbs vs Iliac Branch Devices in Narrow Common Iliac Arteries: A Retrospective Cohort Pilot Study.Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists · 2026
- Management of the Iliac Bifurcation During Endovascular Aneurysm Repair: A Systematic Review and Network Meta-Analysis.European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery · 2026
- A Multicentre Experience of Inner-Branched Endovascular Aneurysm Repair for Treatment of Type Ia Endoleak After EVAR.Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists · 2026
- Modified DrySeal Sheath U-turn Technique with Sheath Escalation and a Triple Coaxial System for Complex Gore Excluder Iliac Branch Endoprosthesis Placement: A Retrospective Cohort.EJVES vascular forum · 2026