The Journal of arthroplasty

Early Aseptic Tibial Loosening Following Cemented and Cementless Fixation of the Same Three-Dimensionally Printed Tibial Baseplate: A Consecutive Cohort of 1,214 Primary Total Knee Arthroplasties.

Alberto Fogacci, Clément Favroul, Cécile Batailler, Elvire Servien, Sébastien Lustig

Department of Orthopaedic Surgery and Sports Medicine, FIFA Medical Centre of Excellence, Croix-Rousse Hospital, Hospices Civils de Lyon, Lyon, France; II Orthopaedic and Traumatologic Clinic, IRCCS Istituto Ortopedico Rizzoli, Alma Mater Studiorum - University of Bologna, Bologna, Italy. Electronic address: albertofogacci4@gmail.com. · Department of Orthopaedic Surgery and Sports Medicine, FIFA Medical Centre of Excellence, Croix-Rousse Hospital, Hospices Civils de Lyon, Lyon, France. Electronic address: clement.favroul@chu-lyon.fr.

Observational

Key points

From the paper's conclusion
  • Early aseptic tibial loosening was uncommon with both cemented and cementless fixation of the same 3D-printed tibial baseplate. There was no difference in early revision-free survivorship detected, but this should not be interpreted as evidence of equivalence. Longer-term follow-up is required.

Abstract

BACKGROUND: Tibial fixation is critical to total knee arthroplasty (TKA) survivorship. Although modern three-dimensional (3D)-printed porous titanium tibial baseplates are designed to improve fixation, implant-specific data on early aseptic loosening remain limited. This study evaluated early aseptic tibial loosening of the same 3D-printed tibial baseplate with cemented and cementless fixation. METHOD: This retrospective single-center study included 1,214 consecutive primary TKAs performed between 2018 and 2025 using a specific tibial baseplate: 437 cemented and 777 cementless. The primary endpoint was revision for aseptic tibial loosening. The secondary outcomes included radiographic findings, postoperative complications, reoperations, and clinical outcomes. Revision-free survivorship was assessed using Kaplan-Meier analysis. RESULTS: At a mean follow-up of 39.4 months (range, 12.0 to 95.1), aseptic tibial loosening requiring revision occurred in four knees (0.3%): one cemented and three cementless (0.2 and 0.4%, respectively), with no difference detected in revision-free survivorship (log-rank P = 0.632). Radiolucent lines (RLLs) were identified in 20 knees (1.6%): 17 were non-progressive and three progressive. All three knees with progressive RLLs underwent revision, while one additional revised knee had non-progressive RLLs. The Knee Society Score (KSS) Clinical improved from 61.7 to 90.5, KSS Functional from 65.3 to 89.5, and knee flexion from 118.5 to 131.5°. Complication rates were low. DISCUSSION: The low incidence of aseptic tibial loosening indicates favorable early fixation of this modern 3D-printed tibial baseplate with either fixation strategy. Progressive RLLs were associated with revision, although one additional case without definitive radiographic evidence of loosening required further diagnostic work-up, highlighting both the value and limitations of serial radiographic assessment. CONCLUSION: Early aseptic tibial loosening was uncommon with both cemented and cementless fixation of the same 3D-printed tibial baseplate. There was no difference in early revision-free survivorship detected, but this should not be interpreted as evidence of equivalence. Longer-term follow-up is required.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers