Performance of hydrogen peroxide gas plasma and vaporized hydrogen peroxide sterilizers under long narrow lumen challenge conditions.
Lei Gao, Rui Mao, Qing Zhao, Jin Li, Ruixue Qin, Xuan Xiong, Lin Wen
Department of Medical Equipment, Sichuan Provincial People's Hospital, school of medicine, University of Electronic Science and Technology of China, Chengdu, China. · Central Sterile Supply Department, Sichuan Provincial People's Hospital, school of medicine, University of Electronic Science and Technology of China, Chengdu, China.
Key points
From the paper's conclusion- Local validation should consider combined lumen length and diameter and include distal biological monitoring for challenging lumen configurations.
Abstract
BACKGROUND: Long narrow lumens remain challenging for low-temperature sterilization because sterilant must reach the distal lumen endpoint. We compared hydrogen peroxide gas plasma and vaporized hydrogen peroxide systems under matched lumen challenge conditions. METHODS: Polytetrafluoroethylene process challenge devices combined lumen lengths of 2, 4, and 8 m with internal diameters of 0.8, 1.0, and 1.5 mm. Each condition was tested in 5 standard cycles per system, yielding 90 biological indicator observations. Success proportions and between-system percentage-point differences are reported with 95% confidence intervals (CIs). RESULTS: Overall success was 33/45 (73.3%; 95% CI 59.0-84.0) for ASP/STERRAD 100NX and 36/45 (80.0%; 95% CI 66.2-89.1) for STERIS V-PRO Max; the difference was 6.7 percentage points (95% CI -10.8 to 23.7). At 8 m, success was 6/15 (40.0%) and 8/15 (53.3%), respectively; the difference was 13.3 points (95% CI -20.2 to 43.1). The 8 m × 0.8 mm condition was most challenging. Chemical and biological indicator results diverged in the vaporized hydrogen peroxide system. DISCUSSION: Distal biological indicator results varied mainly with lumen geometry. CONCLUSIONS: Local validation should consider combined lumen length and diameter and include distal biological monitoring for challenging lumen configurations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Use of machine learning to unlock cycle performance and future parametric release of terminally sterilized medical devices at scale using low temperature gaseous hydrogen peroxide (VH2O2): A Review.Journal of applied microbiology · 2026
- Correlation between Pressure and Cumulative Concentration during Vaporized Hydrogen Peroxide Sterilization.The Journal of hospital infection · 2026
- Development of a functional and disinfectable sewage vent high-efficiency particulate air filter system and its validation using vaporized hydrogen peroxide.Applied and environmental microbiology · 2026
- Material Compatibility of Poly(Ether)urethane 80A With Vaporized Hydrogen Peroxide Sterilization: A Comprehensive Evaluation of Oxidative Stability and BiocompatibilityJournal of Biomedical Materials Research Part A · 2026
- In-Situ Disinfection of Hospital Privacy Curtains Using a Next-Generation Aerosolized Hydrogen Peroxide SystemmedRxiv · 2026
- Hydrogen gas sensor using planar lightwave circuitDiscover Sensors · 2026