Differences in Contrast Sensitivity Measurement With Letters and Gratings.
Assunta Di Vizio, Riccardo Rolandi, Alessandro Duse, Giulia Carlotta Rizzo, Roberta Daini, Marialuisa Martelli, Marco Barbieri, Gabriele Vestri,
Department of Science, Roma Tre University, Rome, Italy. assunta.divizio@uniroma3.it. · Research Centre in Optics and Optometry (COMiB), University of Milano Bicocca, Milan, Italy.
Key points
From the paper's conclusion- CSF outcomes depend on stimulus type when measured on the same investigated LCD system. Grating stimuli are susceptible to ceiling effects, whereas letters demonstrate good reliability and provide clinically useful normative data, supporting their use in routine functional vision assessment.
Abstract
PURPOSE: To compare contrast sensitivity function (CSF) measurements obtained with either Gabor gratings or Sloan letters using the same LCD-based device, to evaluate the reliability of letter-based CSF testing and to establish normative values in a young healthy population. METHODS: A new computerised chart for psychophysical measurements in clinical settings was used (Vision Chart, CSO). Three related investigations were conducted. In Study 1 (n = 44), a randomised crossover design measured monocular CSF with gratings and letters at four spatial frequencies (1.5, 3.0, 6.0 and 12.0 cycles/degree) in dominant and non-dominant eyes. In Study 2 (n = 31), letter-based CSF was measured in the dominant eye across three consecutive trials and reassessed in a second session to evaluate intra- and intersession reliability. A two-way repeated-measures ANOVA evaluated stimulus type and spatial frequency effects in Study 1. Reliability in Study 2 was assessed using repeatability and agreement statistics. In Study 3, normative data were derived by pooling findings from Studies 1 and 2 (dominant eyes n = 75; non-dominant eyes n = 44). RESULTS: In Study 1, CSF measurements were significantly higher for gratings than for letters at all spatial frequencies tested in both eyes (Holm-corrected p < 0.001). Significant main effects of stimulus type and spatial frequency, as well as a significant interaction, were observed (all p < 0.001). Grating measurements clustered near the upper measurable limit, indicating a ceiling effect. In Study 2, letter-based CSF showed good within-session reliability (intraclass correlation coefficient (ICC) 0.69-0.88) and moderate-to-good test-retest reliability (ICC 0.56-0.75). Bland-Altman analysis indicated good repeatability. Normative percentiles for letter-based CSF were established. CONCLUSIONS: CSF outcomes depend on stimulus type when measured on the same investigated LCD system. Grating stimuli are susceptible to ceiling effects, whereas letters demonstrate good reliability and provide clinically useful normative data, supporting their use in routine functional vision assessment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-51: Contrast Sensitivity in Older Adults: Comparing Traditional and Digital Measures in Relation to Global Cognitive PerformanceArchives of Clinical Neuropsychology · 2026
- Evaluation and comparison of three popular computer-based contrast sensitivity measurement techniques.Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) · 2026
- Smartphone-Based Campbell-Robson Contrast Sensitivity Assessment: Repeatability and Systematic Bias Relative to the Freiburg Acuity and Contrast Test.Korean journal of ophthalmology : KJO · 2026
- Color and Contrast Sensitivity Threshold Norms Across Age Decades in Normal SubjectsTechnologies · 2026
- The influence of color filters and induced defocus on contrast sensitivity function.Optometry and vision science : official publication of the American Academy of Optometry · 2026
- Contrast Sensitivity Functions in ON and OFF Visual Pathways in Myopia.Investigative ophthalmology & visual science · 2026