Energy deposition rhythm for optimizing high pulse repetition frequency LIBS signals under scan speed-laser frequency interaction.
Maoyuan Pang, Jinyi Li, Weiye Yu, Yujie Zhang, Xiaomeng Li, Zhenyan Feng, Yu Ding, Xinxin Liu
Jiangsu Key Laboratory of Big Data Analysis Technology, Nanjing University of Information Science and Technology, Nanjing, 210044, China. · Jiangsu Key Laboratory of Big Data Analysis Technology, Nanjing University of Information Science and Technology, Nanjing, 210044, China. dingyu@nuist.edu.cn.
Key points
From the paper's conclusion- The results indicated that, at a fixed EDR, proportional increases in scan speed and laser frequency maintained a constant spatial interval between adjacent laser pulses, thereby reducing the interference caused by pulse-spacing variations in local ablation overlap and related ablation conditions.
Abstract
In microjoule high pulse repetition frequency laser induced breakdown spectroscopy systems, the traditional single-variable parameter optimization method can only analyze the effects of scan speed and laser frequency independently, without considering the interaction effect between these two factors. To address this issue, an optimization strategy based on energy deposition rhythm (EDR) was proposed for the proportional optimization of scan speed and laser frequency. To begin with, the interaction between scan speed and laser frequency was analyzed using two-factor analysis of variance. Based on this, the effects of proportionally increasing scan speed and laser frequency on spectral signals were investigated by maintaining a constant ratio between the two parameters under different EDR conditions, followed by quantitative analysis based on the optimized spectral signals. The results indicated that, at a fixed EDR, proportional increases in scan speed and laser frequency maintained a constant spatial interval between adjacent laser pulses, thereby reducing the interference caused by pulse-spacing variations in local ablation overlap and related ablation conditions. Meanwhile, this proportional increase also resulted in a larger number of laser pulses contributing within the fixed 20 ms integration window owing to the increase in laser frequency, thereby enhancing the accumulated spectral emission while maintaining signal stability. The accumulated spectral intensities of Mg, Mn, and Fe increased by approximately 4-9, 5-10, and 4-10 times, respectively. The most pronounced quantitative improvement was observed under EDR = 1/400, where the R2 values for Mg, Mn, and Fe increased from 0.8826, 0.9516, and 0.9061 to 0.9745, 0.9722, and 0.9778. This study provides a simple and broadly applicable strategy for optimizing high pulse repetition frequency LIBS signals and improving multi-element quantitative analysis performance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Signal enhancement in laser-induced breakdown spectroscopy for improved sensitivity and detection limits: Mechanisms and prospects.The Review of scientific instruments · 2026
- Signal strength in high repetition rate ns LIBS: role of heat accumulation and ablation crater overlap effects for stainless steelJournal of the Optical Society of America B · 2026
- Electric field-enhanced trace hafnium detection in zirconium using 532 nm and 1064 nm LIBS: A comparative study.Talanta · 2026
- A Forensic Perspective of Electric Field Enhanced Signal Sensitivity and Repeatability of Laser Induced Breakdown Spectroscopy (LIBS) for Industrial Grade Al-Si Alloy.Applied spectroscopy · 2026
- Uncertainty-guided optimization of spectral bands for two-color thermometry using laser-induced defect fluorescence in fused silica.Optics express · 2026
- Overcoming the sensitivity-selectivity trade-off in laser-induced breakdown spectroscopy (LIBS) for uranium isotope analysis via joint spectral decomposition of single- and double-pulse spectra.Talanta · 2026