Pulse wave velocity measurement assessing the superficial carotid-femoral distance via smartphone-based light detection and ranging.
Igor N Posokhov
West Kazakhstan Marat Ospanov Medical University, Maresyev Str., 68, 030000, Aktobe, Kazakhstan. igor@posohov.ru.
Key points
From the paper's conclusion- Smartphone-based LiDAR telemetry provides distance measurements and cfPWV estimates that are highly concordant with traditional tape measurements, offering a reliable single-operator alternative for superficial distance acquisition. While these findings confirm technical equivalence and intra-observer reliability, further studies incorporating reference standards and quantitative time-motion analyses are required to confirm the clinical workflow advantages of this method.
Abstract
PURPOSE: Carotid-femoral pulse wave velocity (cfPWV) is the gold standard for non-invasive assessment of aortic stiffness. Tape measurement of the carotid-femoral distance typically requires an assistant to hold the tape while the operator maintains the Doller transducer position. This study assessed the concordance and intra-observer reliability of distance measurements obtained via smartphone-based light detection and ranging (LiDAR) technology compared with traditional tape measurement. METHODS: A cohort of 46 adult participants (mean age, 50.0±12.7 years) underwent Doppler ultrasound evaluation to determine carotid and femoral pulse transit times. The operator measured the superficial carotid-femoral distance using tape and smartphone LiDAR in alternating sequences. Duplicate measurements were taken following a washout period to evaluate intra-observer reliability. RESULTS: LiDAR distance measurements strongly agreed with tape measurements (ICC (2,1) = 0.92, 95% CI 0.81-0.97; mean bias <0.01 m). The resulting cfPWV calculations demonstrated good agreement (ICC (2,1) = 0.94, 95% CI 0.86-0.98; mean bias, 0.03 m/s). Both modalities exhibited high intra-observer reliability (tape ICC (3,1) = 0.98; LiDAR ICC (3,1) = 0.99). The diagnostic classification for elevated cfPWV (> 10 m/s) showed near-perfect concordance between the methods (kappa > 0.98). CONCLUSIONS: Smartphone-based LiDAR telemetry provides distance measurements and cfPWV estimates that are highly concordant with traditional tape measurements, offering a reliable single-operator alternative for superficial distance acquisition. While these findings confirm technical equivalence and intra-observer reliability, further studies incorporating reference standards and quantitative time-motion analyses are required to confirm the clinical workflow advantages of this method.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Reference-Assisted FBG Sensing for Simultaneous Carotid-Radial Pulse-Wave Measurement and PWV Estimation.Sensors (Basel, Switzerland) · 2026
- Noninvasive pulse wave velocity measurement: technological progress, clinical validation, and translation challenges.Physiological measurement · 2026
- Evaluation of cardiovascular risk using smartwatch-measured pulse wave velocity in China: a nationwide cross-sectional study.Frontiers in cardiovascular medicine · 2026
- Spectral inference of arterial stiffness from a single distal velocity waveform.PNAS nexus · 2026
- Multi-Target Simultaneous Measurement of Distance and Velocity With Arbitrary Waveform FMCW LiDAR SystemJournal of Lightwave Technology · 2026
- Derivation of Pulse Wave Velocity from Arterial Waveforms in Children.American journal of physiology. Heart and circulatory physiology · 2026