Pulse wave velocity measurement assessing the superficial carotid-femoral distance via smartphone-based light detection and ranging.

Igor N Posokhov

West Kazakhstan Marat Ospanov Medical University, Maresyev Str., 68, 030000, Aktobe, Kazakhstan. igor@posohov.ru.

Key points

From the paper's conclusion
  • Smartphone-based LiDAR telemetry provides distance measurements and cfPWV estimates that are highly concordant with traditional tape measurements, offering a reliable single-operator alternative for superficial distance acquisition. While these findings confirm technical equivalence and intra-observer reliability, further studies incorporating reference standards and quantitative time-motion analyses are required to confirm the clinical workflow advantages of this method.

Abstract

PURPOSE: Carotid-femoral pulse wave velocity (cfPWV) is the gold standard for non-invasive assessment of aortic stiffness. Tape measurement of the carotid-femoral distance typically requires an assistant to hold the tape while the operator maintains the Doller transducer position. This study assessed the concordance and intra-observer reliability of distance measurements obtained via smartphone-based light detection and ranging (LiDAR) technology compared with traditional tape measurement. METHODS: A cohort of 46 adult participants (mean age, 50.0±12.7 years) underwent Doppler ultrasound evaluation to determine carotid and femoral pulse transit times. The operator measured the superficial carotid-femoral distance using tape and smartphone LiDAR in alternating sequences. Duplicate measurements were taken following a washout period to evaluate intra-observer reliability. RESULTS: LiDAR distance measurements strongly agreed with tape measurements (ICC (2,1) = 0.92, 95% CI 0.81-0.97; mean bias <0.01 m). The resulting cfPWV calculations demonstrated good agreement (ICC (2,1) = 0.94, 95% CI 0.86-0.98; mean bias, 0.03 m/s). Both modalities exhibited high intra-observer reliability (tape ICC (3,1) = 0.98; LiDAR ICC (3,1) = 0.99). The diagnostic classification for elevated cfPWV (> 10 m/s) showed near-perfect concordance between the methods (kappa > 0.98). CONCLUSIONS: Smartphone-based LiDAR telemetry provides distance measurements and cfPWV estimates that are highly concordant with traditional tape measurements, offering a reliable single-operator alternative for superficial distance acquisition. While these findings confirm technical equivalence and intra-observer reliability, further studies incorporating reference standards and quantitative time-motion analyses are required to confirm the clinical workflow advantages of this method.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Pulse wave velocity measurement assessing the superficial carotid-femoral distance via smartphone-based light detection and ranging — Science Skim