Clinical Qualification of Subcutaneous Injection Devices for Monoclonal Antibodies
Beate Bittner, Javier Muñoz, Sabine Frank, Michael Dolton, Renato Ravanello, Markus Dembowski, Aude Clement, W. Richter
Roche (Switzerland) · San-Ei Gen F.F.I (Japan)
Key points
From the paper's conclusion- A narrative review of clinical data from 29 pharmacokinetic comparability studies revealed that independent of the molecule and the injection device, comparable pharmacokinetic and tolerability profiles were achieved following manual and automated injection.
Abstract
Subcutaneous autoinjectors serve as a convenient and user-friendly alternative to manual injection using prefilled syringes or handheld syringes, enhancing self-administration in decentralized care settings, patient compliance, and dosing accuracy. Autoinjectors consistently penetrate the skin to the same depth, and are, thus, expected to reduce variability in injection conditions as compared to more operator-dependent manual self-injection. Even with an established autoinjector platform, molecule-specific technical development requirements often preclude the availability of the automated device at the start of pivotal clinical studies. This necessitates manual injections during these studies and frequently subsequent pharmacokinetic comparability studies to transition to the autoinjector. A proposed Molecule-Independent Device Bridging Approach (MIDBA) aims to eliminate the need for dedicated pharmacokinetic comparability studies for each new monoclonal antibody developed with a previously validated autoinjector platform by referring to PK data from other mAbs with dosing volumes up to 2 mL. The rationale for this approach lies in the slow absorption rates of mAbs following SC injection, which is primarily influenced by the residence time in subcutaneous tissue and subsequent lymphatic drainage rather than the specifics of the injection method. A narrative review of clinical data from 29 pharmacokinetic comparability studies revealed that independent of the molecule and the injection device, comparable pharmacokinetic and tolerability profiles were achieved following manual and automated injection. The review supports the applicability of the MIDBA as a streamlined development pathway for autoinjector validation without molecule-specific pharmacokinetic studies. This framework is expected to present a novel methodology for efficient transition from manual to automated subcutaneous injections for biotherapeutics.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The molecule-independent device bridging approach (MIDBA): perspective on advancing the evidence base for platform-informed bridging.Expert opinion on drug delivery · 2026
- Pharmacokinetic modeling and quantitative prediction of subcutaneous absorption of antibody-based therapeutics – An updateAdvanced Drug Delivery Reviews · 2026
- Immunogenicity of biologics delivered in the subcutaneous space: Delineating the contributing factors.Journal of pharmaceutical sciences · 2026
- Advancing Subcutaneous Delivery: Evaluation of Near-Body Auto-Injector Systems for High-Volume Biologic Administration with Human Hyaluronidase: Poster Presented at PDA Week 2026.PDA journal of pharmaceutical science and technology · 2026
- Current practices, challenges, and future directions in subcutaneous oncology monoclonal antibody delivery: a qualitative studyExpert Opinion on Drug Delivery · 2026
- Patient-Centric Development of a Low-Volume, Low-Viscosity, Once Every 4 Weeks (Q4W) Autoinjector Combination Product for PovetaciceptbioRxiv · 2026