An integrated thin-film microfluidic point-of-care testing system for rapid detection of Ureaplasma urealyticum in cervical swabs.
Jiayu Wang, Keyu Zhu, Minghui Xue, Fei Su, Haiou Yang, Xiaorui Liu
Shanghai Key Laboratory of Embryo Original Disease, Department of Laboratory Medicine, The International Peace Maternity and Child Health Hospital, School of Medicine, Shanghai Jiao Tong University, Shanghai, China. · Department of Research and Development, IngeDx Technologies Co. Ltd., Suzhou, 215163, China.
Key points
From the paper's conclusion- This system may be well suited for primary healthcare, epidemic response, and resource-limited settings, offering an efficient solution to enhance diagnostics and reduce patient waiting time and supporting timely responses to public health needs.
Abstract
OBJECTIVE: Ureaplasma urealyticum (UU) is a common urogenital microorganism associated with a range of genitourinary and perinatal conditions; rapid and reliable detection is valuable in selected clinical contexts. However, traditional bacterial culture and PCR methods are limited by their speed and involve complex operating procedures. METHODS: We evaluated an integrated thin-film microfluidic sample-to-answer system (ingeDx) for nucleic acid-based detection of UU in cervical swab samples. The system integrates cell lysis, nucleic acid purification, amplification and detection, enabling rapid (~ 45 min for the entire process) analysis with Ct values consistent with standard quantitative PCR. RESULTS: The lowest concentration evaluated (200 copies/mL) was detected in 10 of 10 replicates, although a formal limit-of-detection determination was not performed. No cross-reactivity was observed with the 16 other common pathogens tested; the coefficient of variation of Ct values at the evaluated concentration was below 3%, and the cost per test remained below $1. Clinical validation on 113 cervical swab samples demonstrated strong agreement with a reference NMPA-registered real-time PCR kit: positive agreement 98.08% (95% CI: 89.88%-99.66%), negative agreement 95.08% (95% CI: 86.51%-98.31%), overall agreement 96.46% (95% CI: 91.25%-98.61%), and Kappa 0.929 (95% CI: 0.861-0.997); the AUC for UU was above 0.98. CONCLUSIONS: This system may be well suited for primary healthcare, epidemic response, and resource-limited settings, offering an efficient solution to enhance diagnostics and reduce patient waiting time and supporting timely responses to public health needs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Rapid molecular detection, prevalence and antimicrobial resistance of Ureaplasma urealyticum in genital tract specimens: a study at a district general hospital in China.Diagnostic microbiology and infectious disease · 2026
- Performance evaluation and application of a multiplex PCR capillary electrophoresis method for detecting nucleic acids of seven sexually transmitted pathogensFrontiers in Cellular and Infection Microbiology · 2026
- An integrated valved microfluidic platform for rapid and simultaneous nucleic acid detectionLab on a Chip · 2025
- Integrated Microfluidic CRISPR-Based Diagnostic Device for Rapid Testing of HPVs.Biosensors · 2026
- Finger-actuated microfluidic device for point-of-care circulating cell-free nucleic acid extraction.Lab on a chip · 2026
- An Integrated Centrifugal Microfluidic Point-of-Care Low-Cost Chip for Rapid Pathogen Nucleic Acid Detection and Diagnosis of Bloodstream InfectionsAnalytical Chemistry · 2026