Monte Carlo-validated neural network for projection-domain material decomposition in dual-energy CBCT.
Xin Ming, Huipeng Meng, Hezheng Zhai, Chengwen Yang
School of Biomedical Engineering and Technology, Tianjin Medical University, 22 Qixiangtai Road, Heping District, Tianjin, 300070, China. xin.ming@tmu.edu.cn. · Department of Radiation Oncology, School of Medicine, Tianjin First Central Hospital, Nankai University, Tianjin, 300192, China.
Key points
From the paper's conclusion- In conclusion, the proposed MC-validated NN framework provides a robust and accurate approach to DE-CBCT material decomposition.
Abstract
This study aimed to develop and validate a neural network (NN)-based projection-domain material decomposition framework for dual-energy cone-beam CT (DE-CBCT) to improve image quality and quantitative accuracy. A fully connected neural network, trained on a high-fidelity Monte Carlo simulation database of basis material projections, was used for material decomposition. Its performance was benchmarked against a conventional polynomial fitting (POLY) method. The assessment included relative electron density (RED) accuracy, contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) of virtual monoenergetic images (VMIs), and the reduction of metal (MASI) and streak (SAI) artifacts.The NN method demonstrated improved performance in handling extreme-density materials over the POLY method. It reduced RED error from 26.2% to 4.5% for the titanium insert, and from 45.6% to 19.9% for low-density lung insert. NN-derived VMIs showed increased CNR (improvements up to 88.2% for high-Z materials) and higher SNR (average 78.9% improvement). Substantial artifact suppression was observed, with MASI reduced by 34.3% and SAI by 85.1%. In the head phantom, the NN method improved CNR by 12.5-14.0% over single-energy CBCT while reducing noise by 22.3%. In conclusion, the proposed MC-validated NN framework provides a robust and accurate approach to DE-CBCT material decomposition. It produces VMIs with improved contrast, reduced noise, and suppressed artifacts, addressing key limitations of conventional CBCT. This method supports the potential of CBCT for high-precision quantitative applications in image-guided and adaptive radiotherapy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development of a novel material specific imaging algorithm and dual-energy CBCT technology for the next generation imaging solution on O-ring linacs.Physics in medicine and biology · 2026
- PDDUNet: A primal-dual deep unrolling network for CT metal artifact reduction.Journal of X-ray science and technology · 2026
- Revealing hidden material composition via physics-induced deep learning for precision clinical diagnosis.Patterns (New York, N.Y.) · 2026
- Dual-energy cone-beam CT improves iodine contrast and Hounsfield unit value stability in the maxillofacial region: A phantom study.Biomedical physics & engineering express · 2026
- A competition-proven deep learning-based dual-domain framework for low-dose CBCT reconstruction.Quantitative imaging in medicine and surgery · 2026
- Deep learning‐based dual‐domain reconstruction for nonstop gated CBCT in respiratory gating lung SBRTMedical Physics · 2026