Development of the noise theoretical formula for analyzing noise components on CT images.
Keisuke Fujii, Kuniharu Imai
Nagoya University Graduate School of Medicine, 1-1-20 Daiko-Minami, Higashi-ku, Nagoya, 461-8673, Aichi, Japan. fujii@met.nagoya-u.ac.jp. · Nagoya University Graduate School of Medicine, 1-1-20 Daiko-Minami, Higashi-ku, Nagoya, 461-8673, Aichi, Japan.
Key points
From the paper's conclusion- We established a theoretical framework to analyze the individual noise components on computed tomography (CT) images and evaluated the relationship between CT doses and each component.
Abstract
We established a theoretical framework to analyze the individual noise components on computed tomography (CT) images and evaluated the relationship between CT doses and each component. The model was developed based on CT numbers and the relative noise standard deviation (SD) measured directly on the CT images, enabling noise decomposition solely from image-domain characteristics without relying on raw projection data. CT images of a quality control phantom were obtained at various CT doses using two CT scanners. We measured mean CT numbers and the relative noise SD at each rod position on the CT images and estimated each noise component using the proposed framework. The relative noise SD was expressed as a function of the transmitted dose corresponding to the number of photons through the phantom, where the respective terms in the formula represent the contributions of structural, quantum, and electronic noise components. Results showed that the proportion of quantum noise exceeded 70% at CT doses routinely used in clinical settings, remaining the dominant component. However, at lower doses (CTDIvol ≤1 mGy), the contribution of electronic noise became substantial, reaching approximately 10-30%. This framework can serve as a baseline for future CT dose optimization and provide useful insights for developing image reconstruction and processing methods to effectively reduce image noise.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- "Quantitative Modeling and Optimization of Radiation Dose in Routine CT Imaging: Balancing Image Quality and Dose Reduction under the ALARA Principle Framework"International Journal of Drug Delivery Technology · 2026
- Evaluation of Image Noise from Computed Tomography in an Organ-Specific Anthropomorphic PhantomarXiv · 2026
- CT Radiation Dose Reduction With Preserved Diagnostic Performance: How Far Have We Come Over 25 Years?American Journal of Roentgenology · 2026
- Impact of noise power spectrum (NPS) peak frequency on CT radiomics features: A computational phantom study.Biomedical physics & engineering express · 2026
- Development of an Electronic Clinical Quality Measure for Excessive Radiation Dose or Inadequate Image Quality in Diagnostic CT in Adults.Journal of the American College of Radiology : JACR · 2026
- Shapley effects to identify the most influential input parameters of organ dose estimation following CT scans in childhood.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026