Impact of Implantation Site and Bioprosthetic Valve Type on Valve Failure: Analysis of Explanted Valves from Chinese Patients.

Xiaolan Tang, Wentao Feng, Haibo Zhang, Li Li, Guangpu Fan, Hongyue Wang, Xuejing Duan, Wei Zhai,

Key Laboratory of Biomechanics and Mechanobiology (Beihang University), Ministry of Education, Key Laboratory of Innovation and Transformation of Advanced Medical Devices, Ministry of Industry and Information Technology, National Medical Innovation Platform for Industry-Education Integration in Advanced Medical Devices (Interdiscipline of Medicine and Engineering), School of Biological Science and Medical Engineering, Beihang University, Haidian District, No.37 Xueyuan Road, Beijing, 100191, China. · Key Laboratory of Biomechanics and Mechanobiology (Beihang University), Ministry of Education, Key Laboratory of Innovation and Transformation of Advanced Medical Devices, Ministry of Industry and Information Technology, National Medical Innovation Platform for Industry-Education Integration in Advanced Medical Devices (Interdiscipline of Medicine and Engineering), School of Biological Science and Medical Engineering, Beihang University, Haidian District, No.37 Xueyuan Road, Beijing, 100191, China. fengwnt@buaa.edu.cn.

Key points

From the paper's conclusion
  • Biomechanical differences in implantation site and bioprosthetic valve type significantly impact valve failure characteristics, which should be considered in future valve improvements and clinical selection strategies. Moreover, we found that rapidly and gradually failed bioprostheses exhibited distinct failure characteristics; valve durability was not reduced in younger patients.

Abstract

PURPOSE: Analysis of explanted bioprosthetic heart valves reveals causes of valve failure. While explant studies are common, factors driving rapid valve deterioration remain unclear. Moreover, long-term valve durability is biomechanically influenced. Understanding how implantation site and bioprosthetic valve type biomechanically affect failure modes is crucial to optimizing valve design and longevity. METHODS: 121 failed bioprostheses were explanted from Chinese patients aged 5 to 78 years. Patient and valve data were collected. Failure characteristics and rapid-deterioration characteristics were examined and analyzed. Influences of implantation site (aortic vs. mitral) and bioprosthetic valve type (porcine aortic valve vs. bovine pericardial valve) on failure were statistically assessed. RESULTS: Younger patients exhibited comparable or exceeding implantation durations versus older patients (age ≤ 40 at implant: 121 ± 60 months vs. age > 40: 95 ± 46 months, p = 0.019). Compared to the group with bioprosthetic in-vivo durability > 5 years, vegetation significantly contributed to rapid failure (>5 years: 9% vs. ≤ 5 years: 45%, p < 0.001). Porcine aortic valves and valves in mitral sites (higher diastolic pressure) showed susceptibility to commissural dehiscence or tears. Calcification distribution correlated positively with high-stress regions. Pannus overgrowth, influenced by hemodynamics, was more prevalent on inflow surfaces (higher flow velocity) than outflow surfaces (98% vs. 67%, p < 0.001). CONCLUSIONS: Biomechanical differences in implantation site and bioprosthetic valve type significantly impact valve failure characteristics, which should be considered in future valve improvements and clinical selection strategies. Moreover, we found that rapidly and gradually failed bioprostheses exhibited distinct failure characteristics; valve durability was not reduced in younger patients.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Impact of Implantation Site and Bioprosthetic Valve Type on Valve Failure: Analysis of Explanted Valves from Chinese Patients — Science Skim