High-Fidelity Modeling and in Vitro Validation of a Novel NiTi Stent Retriever.
Shuo Wu, Virginia Fregona, Valentina Quartieri, Shengnan Lyu, Francesco Migliavacca, Giulia Luraghi, Tianxiao Zhang, Francesca Berti
Department of Mechanical Engineering and Automation, Robotics Institute, Beihang University, Road Xueyuan 37, Beijing, 100191, China. · Department of Chemistry, Materials and Chemical Engineering, Politecnico di Milano, Piazza L. da Vinci 32, 20133, Milan, Italy.
Key points
From the paper's conclusion- The discretization strategy affected both device-level mechanical prediction and procedure-level retrieval response. The mechanically validated and interface-calibrated beam model provides an efficient framework for simulating NiTi stent retriever procedures in rigid in vitro vessel models. The identified interface coefficient may serve as a reference for similar NiTi-glass retrieval simulations.
Abstract
PURPOSE: Nickel-titanium (NiTi) stent retrievers leverage superelasticity to navigate tortuous vascular anatomies, yet their intricate geometries pose significant challenges for accurate finite element (FE) modeling. This study evaluates the influence of mesh type and design representation on the predictive fidelity of FE simulations for a novel NiTi stent retriever. METHODS: NiTi tube properties for a novel stent retriever prototype were characterized. FE models were developed using tetrahedral and hexahedral solid elements, as well as beam elements with alternative junction representations. Lateral compression, axial tension, and bending simulations were validated against experiments. The selected beam formulation was then incorporated into a full-process model of crimping, delivery, deployment, and retrieval, with in vitro validation in straight, L-shaped, and U-shaped glass vessel models. RESULTS: The hexahedral solid model showed the closest agreement with device-level experiments and most clearly resolved local stress and strain concentrations. Among the beam formulations, the T-type junction model preserved the primary load-transfer path and provided an efficient representation for full-process simulations. After interface calibration, the full-process model reproduced deployment configuration, retrieval-force evolution, and procedure-dependent deformation. In the U-shaped model, the predicted mean retrieval force differed from the experiment by 2.1%, and peak axial strain remained below 8%. CONCLUSION: The discretization strategy affected both device-level mechanical prediction and procedure-level retrieval response. The mechanically validated and interface-calibrated beam model provides an efficient framework for simulating NiTi stent retriever procedures in rigid in vitro vessel models. The identified interface coefficient may serve as a reference for similar NiTi-glass retrieval simulations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Transformation Exclusion Zone and Stress-Induced Superelastic Response in Free-Standing NiTi nanoparticles: A Simulation ApproachPhysica Scripta · 2026
- The ONE File: Design Concept and Finite Element Analysis of a Novel Sixth-Generation 3D-Printed Thermoplastic Polyurethane Hollow Lattice Endodontic Instrument with Combined Rotational-Vibrational Motion.Journal of endodontics · 2026
- Efficient and validated finite element analysis for quantitative assessment of fracture healing.Journal of the mechanical behavior of biomedical materials · 2026
- Study on identifying a consistent metric for comparison of fatigue performance of superelastic Nickel Titanium across different geometries and loading conditions.Journal of the mechanical behavior of biomedical materials · 2026
- Hemodynamic-Structural Multi-Fidelity Assessment of Endovascular Flow Diversion Strategies for Cerebral Aneurysm Treatment: An Integrated Computational and Experimental FrameworkbioRxiv · 2026
- Realistic anatomical model reconstruction and high-fidelity flow computation in stented coronary arteries for the analysis of in-stent thrombosis.Frontiers in cardiovascular medicine · 2026