Clinical journal of gastroenterology

Hypoxemia caused by a body-positioning device during endoscopic retrograde cholangiopancreatography in a patient with severe obesity.

Eisuke Ozawa, Kosuke Takahashi, Junya Tsurifune, Mizuki Kitagawa, Masashi Shibata, Masanori Fukushima, Ryu Sasaki, Masafumi Haraguchi,

Department of Gastroenterology and Hepatology, Nagasaki University Graduate School of Biomedical Sciences, 1-7-1 Sakamoto, Nagasaki, 852-8501, Japan. · Department of Gastroenterology and Hepatology, Nagasaki University Graduate School of Biomedical Sciences, 1-7-1 Sakamoto, Nagasaki, 852-8501, Japan. takapochi0809@gmail.com.

Case report

Key points

From the paper's conclusion
  • We report device-related hypoxemia refractory to airway management that recurred on reapplication.

Abstract

Body-positioning devices are increasingly used during endoscopic retrograde cholangiopancreatography (ERCP). We report device-related hypoxemia refractory to airway management that recurred on reapplication. A 69-year-old woman with class III obesity (body mass index 42.5 kg/m²) underwent prone endoscopic papillectomy under midazolam, pethidine, and dexmedetomidine (DEX); earlier ERCP and endoscopic ultrasonography without the device had been uneventful. With the device applied and hardened before sedation, saturation began to fall during induction, before endoscope insertion, and reached 60%. Jaw thrust and oxygen at 10 L/min did not help; saturation recovered after endoscope withdrawal and device release, and the procedure was abandoned. At a second procedure under lighter sedation without DEX, desaturation to 86% recurred, resisted a nasopharyngeal airway and high-flow nasal oxygen, and resolved on device release alone; the papillectomy was then completed. A retrospective study reported more hypoxemia with this device (14.5% vs. 8.8%), described as mild; our case shows that such hypoxemia can be severe and unresponsive to standard measures. The mechanism was not measured; we suggest that the hardened mat restricts ventilation and airway patency in sedated, severely obese patients. In such patients, hypoxemia unresponsive to airway maneuvers should prompt early device release.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers