Safety of Ovine Forestomach Matrix Across 474 Soft Tissue Defects: Interim Results from the Prospective Multicenter MASTRR Registry.
Tracee Short, John Lawlor, Patrick Martyka, Joshua H Wolf, Jessica M Felton, Alison A Smith, Yosef Y Nasseri, Moshe Barnajian,
Short Consulting Group, 9618 Jefferson Hwy Suite D149, Baton Rouge, LA, 70809, USA. · Associates in Medicine and Surgery, 50 Bellmont St Suite 2, Labelle, FL, 33935, USA.
Key points
From the paper's conclusion- The registry framework provides a valuable mechanism for evaluating device safety and performance under real-world conditions. OFM-based devices demonstrated a favorable safety profile across a broad range of reconstructive applications, surgical specialties, and defect types, with only a single application required in most patients.
Abstract
INTRODUCTION: Well-powered, prospective, registry-based clinical studies present the opportunity for broad safety surveillance of bioscaffolds used in soft tissue reconstruction but remain limited in the field. The Myriad™ Augmented Soft Tissue Reconstruction Registry ("MASTRR Registry") was established as an ongoing multicenter prospective registry to evaluate the real-world safety and performance of ovine forestomach matrix (OFM)-based devices in complex soft tissue reconstruction. METHODS: The interim safety analysis included prospective subject data from the first 411 subjects and 474 defects undergoing reconstruction with OFM-based devices from 10 participating sites. The analysis included subjects from trauma, colorectal, burn, and limb salvage specialties. All adverse events (AEs) were categorized based on severity, causality in relation to the investigational device, and correlated to the surgical procedure and relationship to the index defect. RESULTS: Subjects represented a medically complex population with substantial comorbidity burden and challenging defect characteristics. Most defects were chronic (59.8%) and classified as Centers for Disease Control class II (38.2%) or III (50.4%), indicating that nearly 90% of defects were clean-contaminated or contaminated. Exposed structures (11.4%) and osteomyelitis (15.0%) were also common. A broad range of defect etiologies were included, such as diabetic foot ulcers (22.6%), traumatic injuries (16.5%), ostomy takedowns (11.2%), and pilonidal sinuses (10.8%). Across 411 subjects 12.2% experienced one or more AE. No AEs were determined to be definitively related to OFM-based devices and only a single AE (0.2%) was considered probably device-related. AEs occurring at the index defect were infrequent and included superficial infection (2.9%), dehiscence (3.2%), and deep tissue infection (0.7%). CONCLUSION: The registry framework provides a valuable mechanism for evaluating device safety and performance under real-world conditions. OFM-based devices demonstrated a favorable safety profile across a broad range of reconstructive applications, surgical specialties, and defect types, with only a single application required in most patients. TRIAL REGISTRATION: ClinicalTrials.gov identifier, NCT05243966.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Vascularized tissue coverage of trauma and acute care surgery defects with ovine forestomach matrix: interim results of a prospective multicenter studyJournal of Trauma and Injury · 2026
- Management of volumetric muscle loss in a dog using bioactive glass-alginate-cellulose composite and a transpositional flap: a case report.Veterinary research communications · 2026
- Selective use of reinforced tissue matrix in complex abdominal wall reconstruction: a multicenter consensus analysis.Journal of abdominal wall surgery : JAWS · 2026
- Amniotic membrane for oral surgical defect coverage: A clinical study.Bioinformation · 2026
- Single-layer 2-mm 3-dimensional Bovine Collagen-Elastin Matrix for Reconstruction of Superficial Intraoral Defects After Tumor Resection.Plastic and reconstructive surgery. Global open · 2026
- NovoSorb™ biodegradable temporizing matrix for post-excision wound reconstruction in a paediatric giant congenital melanocytic naevus: a case report.Journal of surgical case reports · 2026