Feasibility and durability of annular stabilization using an internal fixation ring in aortic valve repair: a computed tomography-based study.
Hidefumi Nishida, Kohei Abe, Shunsuke Shibata, Minoru Tabata
Department of Cardiovascular Surgery, St. Luke's International Hospital, 9-1 Akashi-cho, Chuo-ku, Tokyo, 104-0044, Japan. nishida.hidefumi.n2@luke.ac.jp. · Department of Cardiovascular Surgery, St. Luke's International Hospital, 9-1 Akashi-cho, Chuo-ku, Tokyo, 104-0044, Japan.
Key points
From the paper's conclusion- The internal annular fixation ring technique provides feasible and durable aortic annulus stabilization in isolated aortic valve repair. Moderate AI recurrence highlights the need for further technical refinements.
Abstract
OBJECTIVES: Aortic annulus dilatation is a critical risk factor for late aortic insufficiency (AI) recurrence after isolated aortic valve repair. Durable annular stabilization remains technically challenging, and long-term data are limited. We evaluated the mid-term geometric stability of a novel "internal annular fixation ring" technique using an expanded polytetrafluoroethylene (ePTFE) ring and metallic holder. METHODS: We retrospectively analyzed 27 consecutive patients who underwent isolated aortic valve repair with internal annular fixation ring technique for significant AI between January 2015 and April 2024. Aortic root geometry was assessed using cardiac computed tomography (CT) and transthoracic echocardiography, preoperatively and during follow-up. The mean follow-up period for CT was 49.0 ± 38.2 months. The primary outcome was aortic annular stability assessed by agreement between the intended intraoperative annular diameter and the average annular diameter or area-derived annular diameter measured on follow-up CT. Secondary outcomes included freedom from aortic valve reintervention, echocardiographic and CT findings, late survival, and freedom from moderate AI. RESULTS: Overall, concomitant leaflet repair was performed in 74.1% of patients. The intended intraoperative annular diameter ((22 mm (22-24)) was not significantly different from the average annular diameter (22.2 mm (21.4-23.4), p = 0.49) or the area-derived annular diameter (22.2 mm (21.0-23.3), p = 0.15) measured on the follow up CT. Freedom from aortic valve reintervention was 92.3% at 7 years. Although no severe AI recurred, moderate AI developed in 7 patients (25.9%); however, no patient required hospitalization for heart failure during follow-up period. CONCLUSIONS: The internal annular fixation ring technique provides feasible and durable aortic annulus stabilization in isolated aortic valve repair. Moderate AI recurrence highlights the need for further technical refinements.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Annuloplasty Ring Size and Long-term Durability after Degenerative Mitral Valve Repair.The Annals of thoracic surgery · 2026
- Aortic valve repair in 2026 - state of the art: what every cardiologist should know.Cardiology journal · 2026
- Five-year durability of aortic root preservation in type A aortic dissection: A single centre experience in 1,333 patients.Interdisciplinary cardiovascular and thoracic surgery · 2026
- Mitral valve repair using the free-knot technique: A simple, precise, and length-stable method for expanded polytetrafluoroethylene chordal adjustment.JTCVS techniques · 2026
- Y-incision aortic annular enlargement using the arc patch modification: a step-by-step tutorial.Multimedia manual of cardiothoracic surgery : MMCTS · 2026
- Efficacy and Safety of Routine Postdilation in Balloon-Expandable Transcatheter Aortic Valve Implantation Platform: The DUOTAP Randomized Clinical Trial.JAMA cardiology · 2026