SS-OCT-guided dual-arm master-slave robotic system for in vivo subretinal injection in porcine eyes.
Zhangwanyu Wei, Zhaodong Wang, Junqi Chen, Yike Huang, Xieyichang Li, Meng Wang, Gengwen Du, Yang Yang,
Department of Ophthalmology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, 100730, China. · School of Mechanical Engineering and Automation, Beihang University, Beijing, 100191, China.
Key points
From the paper's conclusion- These results support the feasibility and preliminary safety of this platform for subretinal injection and warrant further evaluation for remote bimanual vitreoretinal microsurgery.
Abstract
To evaluate the feasibility, safety, and stability of a novel remote multimodal surgical platform that integrates a dual-arm robotic system with an intraoperative swept-source OCT surgical microscope system for subretinal injection in porcine eyes. Subretinal injection was performed in vivo in Bama miniature pigs (one eye per animal), with six animals assigned to robotic manipulation (RM) and six to manual manipulation (MM). Successful injection was defined as localized subretinal bleb formation confirmed by intraoperative OCT. Outcome measures included procedure time, injection time, first-attempt success rate, bleb morphology, intraoperative complications, postoperative OCT findings, needle-tip tremor amplitude, and histological changes. Subretinal injection was successful in all 12 eyes (100% success rate), with a first-attempt success rate of 5/6 in the RM group and 3/6 in the MM group. Total procedure time was longer in the RM group (901.5 vs. 489.8 s, P = 0.0093), whereas injection time did not differ significantly (median 8.00 vs. 8.50 s, P = 0.6842). Tremor amplitudes were numerically lower in the RM group but did not reach statistical significance (average: 4.48 ± 0.62 vs. 11.06 ± 3.65 μm, P = 0.087; peak: 23.19 ± 2.65 vs. 47.98 ± 18.66 μm, P = 0.146). No severe complications (extensive retinal detachment, uncontrolled bleeding, or lens injury) occurred in either group; minor localized hemorrhage showed a narrower distribution in the RM group than in the MM group. Postoperative OCT confirmed retinal reattachment in all eyes, and histology revealed only localized entry-site changes. This dual-arm robotic platform enabled successful bimanual subretinal injection in all porcine eyes, with comparable injection time, bleb morphology, and short-term tissue safety relative to manual surgery. Despite a longer total procedure time, the system showed a non-significant trend toward reduced needle-tip tremor. These results support the feasibility and preliminary safety of this platform for subretinal injection and warrant further evaluation for remote bimanual vitreoretinal microsurgery.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Autonomous robotic intraocular surgery for targeted retinal injectionsScience Robotics · 2026
- Motion Compensation and Adaptive Force Control via iOCT-FBG Sensor Fusion for Robotic Subretinal Injection.IEEE International Conference on Robotics and Automation : ICRA : [proceedings]. IEEE International Conference on Robotics and Automation · 2026
- Ultra-Long-Distance Robotic-Assisted Subretinal Injection for Submacular Hemorrhage.JAMA ophthalmology · 2026
- Telemanipulated Robot-Assisted Vitreoretinal Surgery: Safety and Effectiveness of the Luca™ System in robotically performed core vitrectomy in ERM or Macular Hole Surgical Treatment Safety and feasibility of Luca™ System.Retina (Philadelphia, Pa.) · 2026
- Bimanual Ophthalmic Trocar Insertion: A Comparative Analysis of Manual vs. Robot-Assisted Performance.RO-MAN ... : the ... IEEE International Symposium on Robot and Human Interactive Communication : proceedings. IEEE International Symposium on Robot and Human Interactive Communication · 2026
- Ultrasound-Led Multimodality Imaging-Guided Robotic Needle Insertion With Respiratory Phase Consistency in Phantom and Animal Models.Ultrasound in medicine & biology · 2026