International journal of computer assisted radiology and surgery

Investigation of gear design parameters for soft-gear-driven self-propelled colonoscopy robot.

Keisuke Osawa, Kaiwen Duan, Akio Ueda, Ryu Nakadate, Jumpei Arata, Yoshihiro Nagao, Tomohiko Akahoshi, Masatoshi Eto,

Department of Mechanical Engineering, Faculty of Engineering, Kyushu University, 744 Motooka Nishi-Ku, Fukuoka, 8190395, Japan. k-osawa@mech.kyushu-u.ac.jp. · Graduate School of Information, Production and Systems, Waseda University, 2-7 Hibikino Wakamatsu-Ku, Kitakyushu, 8080135, Japan.

Key points

From the paper's conclusion
  • These results demonstrate the mechanical feasibility of using soft gears as a propulsion mechanism for self-propelled colonoscopy robots. Although the evaluation was limited to a colonoscopy training model, the proposed design demonstrated stable propulsion in a colonoscopy training model while maintaining compliant contact with the intestinal wall. The simple and compact mechanical structure provides a promising proof of concept for future minimally invasive robotic endoscopy, although further ex vivo and in vivo validation will be required.

Abstract

PURPOSE: Colonoscopy is the most effective method for colorectal cancer screening; however, successful insertion requires a high level of operator skill and may cause patient discomfort or injury. To address these challenges, self-propelled colonoscopy robots that generate propulsion at the distal tip have been actively investigated. In our previous work, we developed a self-propelled colonoscopy robot using a worm-gear mechanism; however, the use of rigid gears posed a potential risk of excessive contact force on the colonic wall. METHODS: In this study, we propose a self-propelled colonoscopy robot enabled by soft gears that combine compliant contact with gear-based power transmission. By reinterpreting the crawler-based propulsion concept as a soft spur gear driven by a worm-gear mechanism, the proposed design allows elastic deformation of the gear teeth under load, providing adaptive traction through compliant deformation of the gear teeth. Gear parameters were optimized using gear-design software, and four soft-gear configurations were fabricated to investigate their propulsion characteristics. RESULTS: The prototypes were fabricated using stereolithography 3D printing and evaluated using a colonoscopy training model. The robot achieved a maximum propulsive speed of 9.6 mm/s and a propulsive force of 1.5 N in straight sections. Gears with larger modules produced higher propulsion performance. In addition, the proposed coaxial double-worm-gear architecture enabled traversal of curved sections of the colon model under manual steering. CONCLUSION: These results demonstrate the mechanical feasibility of using soft gears as a propulsion mechanism for self-propelled colonoscopy robots. Although the evaluation was limited to a colonoscopy training model, the proposed design demonstrated stable propulsion in a colonoscopy training model while maintaining compliant contact with the intestinal wall. The simple and compact mechanical structure provides a promising proof of concept for future minimally invasive robotic endoscopy, although further ex vivo and in vivo validation will be required.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Investigation of gear design parameters for soft-gear-driven self-propelled colonoscopy robot — Science Skim