Clinical outcome prediction of Contour Neurovascular System treatment based on pre-treatment geometric and hemodynamic characteristics.
Fina Gießler, Anna Bernovskis, Franziska Gaidzik, Jana Korte, Philipp Berg, Sönke Peters, Naomi Larsen, Sylvia Saalfeld
Institute for Medical Informatics and Statistics, University Hospital Schleswig-Holstein, Kiel, Germany. giessler@medinfo.uni-kiel.de. · Otto-von-Guericke-University Magdeburg, Research Campus STIMULATE, Magdeburg, Germany.
Key points
From the paper's conclusion- Pre-treatment geometric and hemodynamic characteristics are linked not only to clinical outcome but also to post-treatment CNS shape modifications. By linking pre-operative flow conditions to follow-up device behavior, this study improves the understanding of hemodynamic factors associated with CNS shape modifications and suggests that simulation-based analysis may support future patient-specific treatment planning using exclusively pre-operative data.
Abstract
PURPOSE: The contour neurovascular system (CNS) is increasingly used for endovascular treatment of intracranial aneurysms, but outcome and post-treatment device behavior vary between patients. This study investigated whether pre-treatment geometric and hemodynamic characteristics from image-based blood flow simulations are associated with clinical outcome and follow-up CNS shape modifications. METHODS: Pre-treatment vascular geometries and blood flow simulations were analyzed in 39 CNS-treated cases. Geometric, hemodynamic, and ostium-flow clustering features were extracted, followed by global feature thinning to reduce redundancy. Associations with binary clinical outcome and follow-up shape modifications (SM1-SM3) were assessed using nonparametric univariate statistics with false discovery rate correction. Exploratory outcome discrimination was evaluated using a support vector machine classifier. RESULTS: Aneurysm geometry showed the strongest association with clinical outcome. Smaller dome and ostium areas were associated with positive outcome (Cliff's δ = - 0.57 and - 0.43), while hemodynamic parameters showed weaker but directionally consistent trends, including lower peak outflow velocity, and reduced inflow intensity. Shape-modification patterns differed across deformation classes. SM1 and SM2 showed no FDR-significant features but moderate effects for aneurysm geometry and selected hemodynamic metrics. SM3 showed the strongest associations with oscillatory flow characteristics (δ ≈ 0.56-0.57) and peripheral outflow dominance. Exploratory machine learning showed limited, non-significant outcome discrimination (mean AUC = 0.57), with selected features consistent with univariate findings. CONCLUSION: Pre-treatment geometric and hemodynamic characteristics are linked not only to clinical outcome but also to post-treatment CNS shape modifications. By linking pre-operative flow conditions to follow-up device behavior, this study improves the understanding of hemodynamic factors associated with CNS shape modifications and suggests that simulation-based analysis may support future patient-specific treatment planning using exclusively pre-operative data.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Hemodynamic Impact of Morphological Changes in Intracranial Aneurysms Treated with the Contour Neurovascular System: A Pilot In Silico Study.Brain sciences · 2026
- An integrated machine learning framework for cerebral aneurysm rupture status classification using clinical, morphological, and hemodynamic features.PloS one · 2026
- Local pressure as a dominant hemodynamic driver of wall enhancement in anterior communicating artery aneurysms: a facet-level computational fluid dynamics and vessel wall imaging analysis.Quantitative imaging in medicine and surgery · 2026
- Improving Risk Stratification and Surgical Decision-Making for Unruptured Cerebral Aneurysm: A Proof-of-Concept of Novel Morphological Parameters and Computational Hemodynamics.Brain sciences · 2026
- Hemodynamic-Structural Multi-Fidelity Assessment of Endovascular Flow Diversion Strategies for Cerebral Aneurysm Treatment: An Integrated Computational and Experimental FrameworkbioRxiv · 2026
- Graph and Geometric Deep Learning for Intracranial Aneurysm Geometry and Hemodynamics: A Systematic Review with Implications for Neurovascular Implant Design and Evaluation.Life (Basel, Switzerland) · 2026