Fab-Level LC-MS Assessment of Chain Pairing in Asymmetric Bispecific Antibodies Enabled by IdeS Digestion and Selective Hinge Reduction.
Biao Shen, Tri Nguyen, Hui Xiao, Yang Shen, Ning Li
Analytical and Biological Mass Spectrometry, Tarrytown, NY, 10591-6706, USA. · Bispecific and Antibody Engineering, Regeneron Pharmaceuticals, Inc., Tarrytown, NY, 10591-6706, USA.
Key points
From the paper's conclusion- IdeS digestion combined with selective hinge reduction and controlled solution chemistry (PBS + EDTA) provides a practical, scalable LC-MS framework for routine Fab-level chain pairing assessment to support BsAb development and characterization.
Abstract
PURPOSE: Accurate assessment of heavy/light chain pairing is essential for asymmetric bispecific antibodies (BsAbs), because low-level mispairing (including light-chain swapping) can be difficult to resolve at the intact IgG level. We sought to enable robust Fab-level pairing assessment with a workflow broadly applicable across human IgG subclasses. METHODS: IgG-like BsAbs were digested with FabRICATOR (IdeS) to generate uniform F(ab)₂ and Fc fragments, followed by selective hinge reduction using 2-mercaptoethylamine (2-MEA) to produce discrete Fab fragments. Reduction conditions were optimized to minimize over-reduction and suppress 2-MEA-derived Fab adducts. Fab-level species were evaluated by LC-MS, and an RP-LC method was developed to separate Fc from Fab to reduce spectral interference. RESULTS: Compared with DTT and TCEP, 2-MEA preferentially reduced hinge-region disulfides to efficiently convert F(ab)₂ to Fab under practical conditions (e.g., 10 mM, 37 °C, 30-60 min). A key analytical artifact was formation of 2-MEA-related Fab adducts (Δmass ~ 76 Da) that can suppress or overlap Fab signals. Adduct formation was reduced in PBS relative to Tris buffer (pH 8) and was further suppressed to near-background levels by inclusion of EDTA (5 mM). Under optimized conditions, Fab-level mispairing was detected across multiple BsAb case studies, including low-concentration early-stage samples analyzed without pre-treatment, and the workflow was demonstrated for IgG1/IgG4-based BsAbs and bispecific modalities (altibodies), as well as IgG2/IgG3 antibodies. CONCLUSIONS: IdeS digestion combined with selective hinge reduction and controlled solution chemistry (PBS + EDTA) provides a practical, scalable LC-MS framework for routine Fab-level chain pairing assessment to support BsAb development and characterization.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Parallel Sequencing of the Two Arms on a Bispecific Antibody by Electron Capture Dissociation on a timsOmni Platform Reveals Several By-products in the Controlled Fab-arm Exchange ProcessAnalytical Chemistry · 2026
- Multimodal chromatography solutions for the removal of light chain impurity from a fab therapeutic.Biotechnology progress · 2026
- Electrostatic inversion in pH-gradient antigen affinity chromatography: The selectivity bias for deamidated trastuzumab proteoforms.Journal of chromatography. A · 2026
- Electron activated dissociation technique for middle-down mass spectrometry analysis of disulphide bonds and free thiols: Application in characterising monoclonal antibody biotherapeutic fragments.Journal of pharmaceutical and biomedical analysis · 2026
- Mass Spectrometric Characterization of Native Insulin Hexamers and Non‐Native Heptamers: Formation and Stability Across Insulin AnalogsJournal of Mass Spectrometry · 2026
- Proximity directs microbial transglutaminase site selectivity in native antibody modificationbioRxiv · 2026