A novel spherical applicator for ultrasound-guided assessment of tumor resection in intracranial surgery.

Jeremiah Scarborough, Brooke Maxwell, Berlin Beaver, Ramon A Siochi, Joshua D Hack, Christopher P Cifarelli

Department of Radiation Oncology, West Virginia University, Morgantown, WV, USA. · School of Science, Technology, Engineering, and Math (STEM), American Military University, Charles Town, WV, USA.

Key points

From the paper's conclusion
  • This phantom-based feasibility study demonstrates that a spherical ultrasound applicator can provide a reproducible acoustic interface for imaging simulated brain tumor resection cavities. The device may facilitate intraoperative assessment of cavity margins and potentially complement intracavitary radiation delivery. Further in vivo evaluation is required to establish clinical utility.

Abstract

PURPOSE: Intraoperative ultrasound (iUS) provides real-time imaging during brain tumor resection, but its utility may be limited by surgical cavity collapse and distortion of surrounding anatomy. We developed a spherical ultrasound applicator intended to provide a reproducible imaging interface within the resection cavity, with the potential to support cavity geometry during post-resection assessment. This study evaluated the feasibility of applicator placement and ultrasound image acquisition across simulated resection cavity geometries and imaging conditions. METHODS: A spherical applicator with a 17.5-mm radius and central ultrasound probe channel was fabricated and evaluated using silicone phantoms representing shallow, deep, and irregular resection cavities. Ultrasound imaging was performed using phosphate-buffered saline (PBS) and 1% albumin as acoustic coupling media. Embedded echogenic targets at depths of approximately 3 and 5 cm were evaluated using contrast-to-noise ratios (CNR). Computed tomography was used to verify phantom geometry and applicator fit. RESULTS: The applicator maintained stable contact with simulated cavity surfaces and enabled visualization of cavity margins across all three phantom geometries. Echogenic targets remained detectable at depths up to 5 cm using either coupling medium. At 3 cm, CNR values were 11.88 with PBS and 11.37 with albumin, decreasing to 6.76 and 6.24, respectively, at 5 cm. Imaging depth, rather than coupling medium, was the predominant determinant of target contrast. Computed tomography confirmed phantom geometry and applicator alignment. CONCLUSION: This phantom-based feasibility study demonstrates that a spherical ultrasound applicator can provide a reproducible acoustic interface for imaging simulated brain tumor resection cavities. The device may facilitate intraoperative assessment of cavity margins and potentially complement intracavitary radiation delivery. Further in vivo evaluation is required to establish clinical utility.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A novel spherical applicator for ultrasound-guided assessment of tumor resection in intracranial surgery — Science Skim