Experimental comparisons of optical coherence tomography-based versus angiography-based time-averaged wall shear stress estimations
Jarka Naser, Nicholas A. Fogell, Miten Patel, Pan Yang, Mayurey Kalaravy, Fotios Savvopoulos, Rob Krams, Jean‐Paul Aben,
Imperial College London · Queen Mary University of London
Key points
From the paper's conclusion- This study suggests that 3D-QCA-based TAWSS is feasible in both normal and stenotic arteries and that further clinical evaluation of rapid TAWSS from 3D-QCA is warranted, which may facilitate clinical adoption of TAWSS assessment.
Abstract
An approach to rapid simulation of time-averaged wall shear stress (TAWSS) on 3D geometries created from 3D Quantitative Coronary Angiography (3D-QCA) methodology has been developed, which enables rapid computational fluid dynamic (CFD) shear stress simulation. We compared TAWSS estimated from 3D-QCA-CFD with optical coherence tomography (OCT)-based CFD simulations in coronary arteries. 15 normal and 5 stenotic coronary arteries in instrumented minipigs were studied. 3D arterial geometries were reconstructed from 3D-QCA and OCT using common centrelines and matched axial positions. Identical boundary conditions were used for both methods through directly measured vessel-specific inlet blood velocities. TAWSS was calculated for axially matched segments (n = 80 for normal arteries; n = 160 for stenotic arteries) and in 3 mm/60° sectors. Mean TAWSS simulation times for 3D-QCA and OCT-based CFD were 17.8 min and ~ 1.5 h respectively. There were significant but numerically small differences in TAWSS for normal arteries (-0.21 ± 0.64 Pa [95%CI -1.04,1.46], p < 0.001), and no significant difference for stenotic arteries (-0.39 ± 3.04 Pa [95%CI -6.35, 5.56], p = 0.25). Axial TAWSS profiles along vessel lengths were similar between the two methods. There is a trend of underestimation by 3D-QCA at higher values of TAWSS compared with OCT, due to differences in geometry dimensions. Similar spatial distributions of TAWSS in both normal and stenotic arteries were observed from co-registered TAWSS maps. This study suggests that 3D-QCA-based TAWSS is feasible in both normal and stenotic arteries and that further clinical evaluation of rapid TAWSS from 3D-QCA is warranted, which may facilitate clinical adoption of TAWSS assessment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Incremental value of coronary computed tomography angiography-derived wall shear stress over stenosis grade in detecting significant coronary artery diseaseEuropean Heart Journal · 2026
- Geometry-Dependent Wall Shear Stress in Standardized 50% Coronary Stenoses: Effects of Pressure and Hematocrit in a Parametric CFD StudyMathematical and Computational Applications · 2026
- Realistic anatomical model reconstruction and high-fidelity flow computation in stented coronary arteries for the analysis of in-stent thrombosis.Frontiers in cardiovascular medicine · 2026
- Computational hemodynamic analysis of idealized coronary arteries with cylindrical and conical stentsComputer Methods and Programs in Biomedicine · 2026
- The influence of stenosis severity and geometric orientation on atherosclerotic progression: A fluid-structure interaction study in the carotid artery.Physical and engineering sciences in medicine · 2026
- Feasibility study for 3D quantitative angiography in internal carotid aneurysms using in silico biplane imaging and 3D vascular geometry constraints.Medical image analysis · 2026