Model-Based Elastic-Resistive Partitioning of Inspiratory Mechanical Power During Conventional and Energy-Targeted Ventilation in a Hybrid Respiratory Simulator.
Barbara Stankiewicz, Marek Darowski, Raman Pasledni, Krzysztof Jakub Pałko, Maciej Kozarski, Krzysztof Zieliński, Tomasz Urbankowski
Department of Modeling and Supporting of Internal Organs Functions, Nalecz Institute of Biocybernetics and Biomedical Engineering, Polish Academy of Sciences, ul. Ks. Trojdena 4, 02-109, Warsaw, Poland. · Department of Modeling and Supporting of Internal Organs Functions, Nalecz Institute of Biocybernetics and Biomedical Engineering, Polish Academy of Sciences, ul. Ks. Trojdena 4, 02-109, Warsaw, Poland. turbankowski@ibib.waw.pl.
Key points
From the paper's conclusion- This study applied a model-based framework to characterize elastic and resistive inspiratory mechanical power during conventional and investigational energy-targeted ventilation in a hybrid respiratory simulator.
Abstract
PURPOSE: Mechanical power provides a global description of the energetic load imposed by mechanical ventilation, but similar total values may reflect different elastic and resistive components. This study applied a model-based framework to characterize elastic and resistive inspiratory mechanical power during conventional and investigational energy-targeted ventilation in a hybrid respiratory simulator. METHODS: An experimental bench study was performed using a hybrid respiratory simulator coupled with a prototype ventilator. Four respiratory system conditions were simulated: normal mechanics and mild, moderate, and severe ARDS-like elastic impairment. Compliance was reduced from 60 to 30 mL/cmH2O, while resistance was kept constant at 10 cmH2O/L/s. Volume-controlled, pressure-controlled, and mechanical power-controlled ventilation were evaluated at respiratory rates of 10, 15, and 20 breaths/min. Maximal and mean elastic and resistive inspiratory mechanical power were assessed. RESULTS: Ventilation mode significantly affected all analyzed power components. Mechanical power-controlled ventilation produced the lowest pooled maximal elastic power, with reductions of 16.2% versus volume-controlled ventilation and 24.3% versus pressure-controlled ventilation (both p < 0.001). This effect was most evident at lower respiratory rates and under more severe elastic impairment. For resistive power, pressure-controlled ventilation produced the highest values, volume-controlled ventilation the lowest, and mechanical power-controlled ventilation remained intermediate. CONCLUSION: Component-level analysis revealed ventilation mode-dependent differences in inspiratory energy transfer that were not captured by total mechanical power alone.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of Airway-Based and Transpulmonary Mechanical Power Across Ventilation Conditions in Moderate-to-Severe ARDS.Life (Basel, Switzerland) · 2026
- Respiratory pyramid interface (RPI): a conceptual three-dimensional pyramid model for visualization of lung mechanics during mechanical ventilation.Frontiers in physiology · 2026
- Discrepancies between estimated and geometrically derived mechanical power in invasively ventilated patients - insights from a randomized clinical trial.Intensive care medicine experimental · 2026
- Flow-controlled ventilation versus pressure-controlled ventilation in moderate to severe ARDS patients: a randomized crossover physiological studyIntensive Care Medicine Experimental · 2025
- Mechanical Power and the Association with Postoperative Impaired Oxygenation and Pulmonary Complications in Orthopedic Patients: Post Hoc Analysis of a Cluster Factorial Randomized TrialAnesthesiology · 2025
- Association Between Mechanical Power and Ventilator-Associated Conditions in Patients Requiring Prolonged Mechanical Ventilation: A Multi-Database Retrospective Cohort Study.Journal of clinical medicine · 2026