Development of a smartphone-based retinal imaging device for stable fundus imaging.
Hidetoshi Maeda, Yoko Fukushima, Kakeru Sasaki, Yoshiyuki Kitaguchi, So Goto, Kohji Nishida
Department of Ophthalmology, Graduate School of Medicine, The University of Osaka, Yamadaoka, Suita, Osaka, 565-0871, Japan. · Department of Ophthalmology, Graduate School of Medicine, The University of Osaka, Yamadaoka, Suita, Osaka, 565-0871, Japan. youko.fukushima@ophthal.med.osaka-u.ac.jp.
Key points
From the paper's conclusion- The device demonstrates significantly improved stability and ease of use compared with conventional methods. Its design may help reduce operator dependency and facilitate fundus image acquisition in supine patients. Further validation in the intended clinical populations is warranted.
Abstract
PURPOSE: To develop a smartphone-based imaging device and evaluate the operability and image acquisition performance. STUDY DESIGN: Prospective pilot study. METHODS: A prototype device was developed, integrating a smartphone holder, a 28-diopter lens holder, and a neck strap. Sixteen examiners (nine ophthalmology residents and seven certified orthoptists) recorded the fundus videos of a healthy volunteer for 60 seconds using three methods: manual handheld, a commercially available non-medical smartphone holder, and our device. Each video was converted into 1,800 frames. Frames were judged as success if they met four criteria: inclusion of the optic disc and macula, fundus coverage > 75%, minimal artifacts, and sharp focus. The image acquisition rate (stability) and the shortest image acquisition time (usability) were compared among the three conditions. RESULTS: The image acquisition rate was significantly higher with our device (31.49 ± 32.53%) compared to the manual handheld (5.31 ± 12.35%; P = 0.0024) and the commercial non-medical holder (3.68 ± 6.96%; P = 0.0057). The shortest image acquisition time was also significantly lower for our device (15.27 ± 24.03 s) than for the other two methods (handheld: 47.10 ± 21.44 s, P = 0.0078; commercial non-medical holder: 36.69 ± 27.29 s, P = 0.0080). CONCLUSION: The device demonstrates significantly improved stability and ease of use compared with conventional methods. Its design may help reduce operator dependency and facilitate fundus image acquisition in supine patients. Further validation in the intended clinical populations is warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Smartphone-Based Fundus Imaging in Global Eye Care: Closing the Gap.Progress in retinal and eye research · 2026
- Digital Anterior Segment Photography in Clinical Practice: Mixed Methods Study.JMIR formative research · 2026
- Comparison of a Smartphone-Based Method for Measuring Anterior Chamber Depth with Anterior Segment OCT and Smith's Technique.Journal of clinical medicine · 2026
- A novel auxiliary method combining a smartphone with a handheld slit-lamp for ocular image documentation of cataracts and diabetic retinopathy in remote regions.BMC ophthalmology · 2026
- Portable nonmydriatic ultra-widefield multispectral fundus camera for comprehensive posterior segment assessment.Biomedical optics express · 2026
- Intensity Variance-Guided Automated Robotic Optical Coherence Tomography System for Ophthalmic Applications.Journal of biophotonics · 2026