A Deep Learning Framework for Retrospective Motion Detection and Conditional Correction (MDCC): Application to Multi-Gradient-Recalled Echo Quantitative MRI.
Gulfam Ahmed Saju, Jeramy Lewis, Satya V V N Kothapalli, Cihat Eldeniz, Dmitriy A Yablonskiy
Mallinckrodt Institute of Radiology, Washington University School of Medicine in St. Louis, 4525 Scott Avenue, St. Louis, MO, 63110, USA. gsaju@wustl.edu. · Mallinckrodt Institute of Radiology, Washington University School of Medicine in St. Louis, 4525 Scott Avenue, St. Louis, MO, 63110, USA.
Key points
From the paper's conclusion- We propose Motion Detection and Conditional Correction (MDCC), a retrospective image-domain pipeline that couples a complex-input volume-level Multi-Echo Motion Detection (ME-MoD) model and a complex-valued Multi-Echo Motion Corrector (ME-MoC) in an iterative detect–correct–re-evaluate loop, where correction is invoked only on volumes flagged as motion corrupted.
Abstract
Multi-echo gradient-recalled-echo (mGRE) MRI supports quantitative neuroimaging by enabling voxel-wise T2*/R2*/R2t* mapping for quantitative assessment of tissue damage and magnetic susceptibility imaging. Subject motion degrades these acquisitions by introducing inconsistencies due to phase-encoding steps collected at different head positions. Although the gradient echoes associated with a given phase-encoding line are acquired at practically the same head position, differences in contrast, signal-to-noise ratio, and B0 field-dependent phase evolution cause the resulting artifacts to appear differently across echoes and can bias downstream parametric maps. Existing learning-based retrospective correction approaches often discard complex phase information, treat multiple gradient-recalled echoes independently, and apply correction indiscriminately to both motion-corrupted and clean data. We propose Motion Detection and Conditional Correction (MDCC), a retrospective image-domain pipeline that couples a complex-input volume-level Multi-Echo Motion Detection (ME-MoD) model and a complex-valued Multi-Echo Motion Corrector (ME-MoC) in an iterative detect-correct-re-evaluate loop, where correction is invoked only on volumes flagged as motion corrupted. ME-MoD reweights each echo with a phase-preserving paired complex gate, followed by a shared residual slice encoder, and a selected volume-level aggregator. ME-MoC is a fully complex-valued, factorized U-Net with echo-axis multi-head self-attention at the bottleneck, trained with a T2*-consistency physics loss. ME-MoD attained an AUC of 0.95 and sensitivity of 0.91 on the held-out detection set. At the cohort-level analysis, ME-MoC improved downstream R2* fitting results relative to the motion-corrupted input and 2D U-Net and achieved the lowest numerical R2* root-mean-square error (RMSE) among the evaluated methods, including modest numerical improvement over baseline.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multi-Stage Phase-Aware Self-Supervised Reconstruction for Multi-Echo Gradient-Echo MRIarXiv · 2026
- A physics‐informed deep learning model for MRI brain motion correctionMedical Physics · 2025
- Robust T1ρ , T2 , and T2 * mapping via spin-locked MOLED with synthetic data-driven deep learning reconstruction.Physics in medicine and biology · 2026
- Training-Free Motion-Corrected 3D Brain MRI Reconstruction via Event-Guided ReferencesarXiv · 2026
- Unified Motion Correction and Reconstruction for Accelerated Zero Echo Time MRIarXiv · 2026
- Frame-wise multi-echo distortion correction for superior functional MRIImaging Neuroscience · 2026