Transducin: An Open-Source Pipeline Recovering SNOMED-CT Coded Measurements from the Undocumented Optopol .OPT and Zeiss Cirrus Private-Tag Formats as DICOM Structured Reports.
Jesús Noel Jaurrieta-Hinojos, José Luis Palomares-Ordóñez, José Francisco Chacón-Hinojos, Miguel Ángel Folgueras-Batres
Retina and Vitreous Service, Hospital Ángeles Chihuahua, Chihuahua, México. noel@oftalmos.org. · Oftalmos.ORG Project, Chihuahua, México.
Key points
From the paper's conclusion- We present Transducin, an open-source Python library that reverse-engineers two undocumented proprietary OCT formats (the Optopol REVO FC130/REVO 60 .OPT binary format and the Zeiss Cirrus HD-OCT 500 private DICOM tags) through systematic binary analysis cross-referenced against DICOM-exported metadata from the same studies, and converts the recovered measurements into standard, vendor-neutral DIC
Abstract
Quantitative optical coherence tomography (OCT) measurements are essential for retinal disease monitoring, yet leading vendors store acquisition data in undocumented proprietary formats or encode measurements exclusively in private DICOM tags inaccessible to open systems, obstructing interoperability between devices and standards-based clinical imaging infrastructure. We present Transducin, an open-source Python library that reverse-engineers two undocumented proprietary OCT formats (the Optopol REVO FC130/REVO 60 .OPT binary format and the Zeiss Cirrus HD-OCT 500 private DICOM tags) through systematic binary analysis cross-referenced against DICOM-exported metadata from the same studies, and converts the recovered measurements into standard, vendor-neutral DICOM TID 1500 Structured Reports with SNOMED-CT coded findings. Format stability was validated across SOCT software versions 11.5.0-21.5.0, spanning approximately 8 years of development; a laterality-inference hypothesis derived from the device's scan-geometry parameters was tested against known eye assignment; and all generated DICOM objects were checked against pydicom's conformance validator and ingested into a clinical PACS. The validation corpus comprised 475 Optopol .OPT files across two clinical sites: 387 at Site A (Chihuahua; 73 patients) and 88 at Site B (Querétaro), spanning three device models and four SOCT software versions with 100% parse success across nine acquisition types. The Zeiss Cirrus pipeline generated TID 1500 SRs for all 41 applicable studies (100%), yielding CMT 203-630 µm and RNFL 53-123 µm across a clinically representative range. A scan-geometry parameter recorded by the device (OCTPARAMS tag 23, the horizontal position of the foveal or disc reflex relative to scan center) was found to encode ocular laterality through its arithmetic sign-an anatomical consequence of the fovea's nasal position relative to the optic disc-and this laterality-inference method was validated across 18 files from two device models and two software versions with 100% accuracy, requiring no operator data entry or filename convention. Transducin provides the first publicly documented specification of the Optopol .OPT format and the first open-source multivendor pipeline generating SNOMED-CT coded DICOM Structured Reports from both Optopol REVO and Zeiss Cirrus devices, closing a gap explicitly confirmed by both manufacturers' own documentation and final users experience. The code is available at https://github.com/oftalmos-org/transducin (Apache License 2.0).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- import_pentacam: Open-Source OCR Recovery of Corneal Tomography Measurements From Pentacam DICOM, PDF, and Image Archives.Translational vision science & technology · 2026
- Pan-retinal pathology detection in oct scans integrating natural language synthesis with diagnostic annotation.NPJ digital medicine · 2026
- Impact of Optical Coherence Tomography Angiography Scan Protocols on Quantitative Microvascular Biomarkers in Diabetic Retinopathy.Translational vision science & technology · 2026
- Differentiable Inverse Rendering Recovers Intrinsic Properties from Intravascular Optical Coherence TomographyarXiv · 2026
- Effect of Intensity Discretization on Radiomic Features in En Face EDI-OCT of Choroidal NevimedRxiv · 2026
- Cross-Device Field-of-View Scale Differences in Clinical OCT Angiography: Phantom Calibration and Large-Scale Clinical Validation.medRxiv : the preprint server for health sciences · 2026