Networked Radiology Imaging and Treatment Planning Systems and FDA Cybersecurity Regulation: A Difference-in-Differences Analysis of 510(k) Clearance Delays.
Chuelwon Lee
RAQA Team, HUINNO Co., Ltd., 3F, 19 Apgujeong-Ro 79-Gil, Gangnam-Gu, Seoul, 06011, Republic of Korea. lcw@huinno.com.
Key points
From the paper's conclusion- The observed clearance time increase for AI-enabled submissions is associated with, and concentrated in, networked radiology panel imaging and treatment planning systems rather than AI-enabled software broadly, and it accumulates over roughly 6 to 24 months rather than appearing immediately.
Abstract
The objective of this study is to determine whether 510(k) clearance duration for AI-enabled devices changed differentially relative to the general market around October 1, 2023-the date FDA began Refuse-to-Accept enforcement of Section 524B cybersecurity documentation completeness-using a freshly re-extracted, complete FDA 510(k) dataset. We analyzed 83,675 FDA 510(k) records (De Novo excluded; decisions dated January 2000 through August 2, 2026). AI-enabled codes were identified via FDA's officially published AI-Enabled Medical Device List (26 codes, N = 7070). Log-transformed clearance duration was modeled with a difference-in-differences (DiD) regression (product code-clustered standard errors, adjusted for submission type and review pathway); a half-year event study specification tested parallel pre-rule trends and timing, and Bonferroni-corrected Mann-Whitney tests (15 comparisons) assessed code-level heterogeneity; AI-enabled status was assigned at the product code level and was not confirmed at the individual-submission level. After adjustment, the overall market showed no significant post-rule change (Post coefficient = 0.029, p = 0.141), while the AI-enabled cohort showed a robust increase (AI × Post = 0.212, p < 0.001; ≈24% longer). The two half-years nearest the cutoff showed no significant pre-rule trend, though the most distant pre-rule half-year (six half-years prior) showed a significant negative coefficient, an anomaly discussed in Limitations; the AI x Post effect itself became significant approximately 24 months post-implementation. Five codes-diagnostic ultrasound (IYN), CT scanners (JAK), radiology image processing software (LLZ), angiography/fluoroscopy systems (OWB), and radiotherapy-planning software (MUJ)-were Bonferroni significant (all p < 0.001) and together drove the aggregate effect; all five are radiology panel codes. The highest-volume AI code (QIH) showed no significant change after correction. The observed clearance time increase for AI-enabled submissions is associated with, and concentrated in, networked radiology panel imaging and treatment planning systems rather than AI-enabled software broadly, and it accumulates over roughly 6 to 24 months rather than appearing immediately. Because AI-enabled status is defined at the product code level, submissions in the control group may include non-AI cyber devices also subject to the cybersecurity rule, and the concurrent mandatory eSTAR transition is not separately identified; these findings should therefore be read as an association warranting device category-specific, rather than blanket AI/ML, regulatory attention, not as established evidence that Section 524B specifically caused the observed delay.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Regulatory time-lag between CE-marking and FDA approval of artificial intelligence softwaresEuropean Heart Journal · 2026
- FDA-Regulated AI-Enabled Medical Devices With Pediatric IndicationsJAMA Network Open · 2026
- 1,357 AI medical devices cleared, 3 actually tested on patient outcomes.PLOS digital health · 2026
- Machine Learning-Enabled Medical Devices Authorized by the US Food and Drug Administration in 2024: Regulatory Characteristics, Predicate Lineage, and Transparency ReportingBiomedicines · 2025
- Use and Public Reporting of Predetermined Change Control Plans for AI-Enabled Medical DevicesJAMA Health Forum · 2026
- Growth and Structure of the FDA-Authorized Radiology AI Landscape, 1998-2025.Academic radiology · 2026