Impella purge solutions in patients receiving extracorporeal membrane oxygenation.
Nathaniel J Martin, Christoph G S Nabzdyk, Scott D Nei, Andrew N Rosenbaum, Troy G Seelhammer, Patrick M Wieruszewski
Department of Pharmacy, Mayo Clinic, 200 First Street SW, Rochester, MN, 55905, USA. · Department of Anesthesiology, Mayo Clinic, Rochester, MN, USA.
Key points
From the paper's conclusion- We aimed to assess Impella device function in patients supported by ECPELLA receiving heparin-containing or anticoagulant-free (5% dextrose or sodium bicarbonate) purge solutions.
Abstract
The optimal Impella purge solution in patients receiving concomitant ECMO and Impella (ECPELLA) support remains to be elucidated. We aimed to assess Impella device function in patients supported by ECPELLA receiving heparin-containing or anticoagulant-free (5% dextrose or sodium bicarbonate) purge solutions. Of 25 included patients, 14 received anticoagulant-free and 11 received heparin-containing purge solutions, with a median of five days of ECPELLA support. The only episode of Impella device malfunction occurred in the heparin-containing group. Given the low event rate and apparent maintenance of acceptable device longevity, use of anticoagulant-free purge solutions alongside systemic anticoagulation in ECPELLA patients may be reasonable but requires validation in an appropriately powered study.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effects of Extracorporeal Membrane Oxygenation Circuits on Drug Sequestration: A Review of Ex Vivo ExperimentsJournal of Clinical Medicine · 2025
- Challenges and opportunities in ECMO monitoring: A mixed-methods appraisal.The journal of extra-corporeal technology · 2026
- Anticoagulation in Mechanical Circulatory Support: Contemporary Challenges and Future Directions in Cardiovascular Pharmacotherapy.Cardiology and therapy · 2026
- Association Between End-Tidal Carbon Dioxide and Pulmonary Artery Pulse Pressure During Veno-Arterial Extracorporeal Membrane Oxygenation with Impella®: A Retrospective Exploratory Study.Journal of clinical medicine · 2026
- Comparative Performance of Oxygenators for Extracorporeal Membrane Oxygenation in Adults: A Retrospective Analysis.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- Comparing hemolysis among extracorporeal pumps: A large-scale retrospective study.Perfusion · 2026