Visualization of instruments-endoscope interference in neuroendoscopic surgery.
Kento Takahara, Masahiro Ogino, Ryo Ueda, Tomoru Miwa, Masaaki Nishimoto, Nobuhiro Yamada, Masahiro Toda
Department of Neurosurgery, Ashikaga Red Cross Hospital, 284-1 Yobe-Cho, Ashikaga-Shi, Tochigi, 326-0843, Japan. pami529@yahoo.co.jp. · Department of Neurosurgery, Ashikaga Red Cross Hospital, 284-1 Yobe-Cho, Ashikaga-Shi, Tochigi, 326-0843, Japan.
Key points
From the paper's conclusion- RZ was successfully visualized and quantitatively characterized. Instrument-endoscope interference is largely governed by the positional relationship of instruments and the endoscope at the entry and by their appearance within the endoscopic field of view. Strategic control of these factors may therefore help minimize interference.
Abstract
PURPOSE: In neuroendoscopic surgeries, instrument-endoscope interference can prevent instruments from reaching certain areas visible within the endoscopic view, leading to the perception of a "restricted zone" (RZ). We aimed to visualize and quantitatively evaluate the RZ, thereby enabling intuitive understanding of instrument-endoscope interference. METHODS: Cadaver studies were conducted to investigate the typical endoscope-target distance, a key factor affecting RZ formation. Based on this distance, a simulated surgical field was created using an endoscope and light source, with light projected from the instrument's entry point. Illuminated regions represented areas accessible to a straight instrument, while shadows cast by the endoscope or other instruments indicated unreachable areas-RZs-visualized within the endoscopic view. The study examined changes in the RZs under various conditions. RESULTS: In cadaver experiments, the typical endoscope-target distance was approximately 7-20 mm. The visualized RZ consistently appeared on the side opposite the instrument in the endoscopic view regardless of endoscopic tip movement. Under typical conditions, the RZ occupied 6.0-28.2% of the endoscopic field of view. Oblique scope or thin scope altered RZ. In simulated three-hand surgery, additional RZs generated by instruments resulted in a more complex RZ pattern, according to the entry-site configuration of the endoscope and the two instruments. CONCLUSIONS: RZ was successfully visualized and quantitatively characterized. Instrument-endoscope interference is largely governed by the positional relationship of instruments and the endoscope at the entry and by their appearance within the endoscopic field of view. Strategic control of these factors may therefore help minimize interference.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development and Evaluation of a Dual-Articulating Rigid Endoscope for Neuroendoscopic Surgery.World neurosurgery · 2026
- Drape-Compatible Tool-Tip Localization for Hand-Held Laparoscopic Instruments via UWB Carrier-Phase Ranging and Trocar-Constrained GeometryarXiv · 2026
- InteractOR: Interacting with Images during Surgery in Mixed Reality Through Real-Time Instrument Segmentation.Healthcare technology letters · 2026
- Impact of localized diminished illumination for needle targeting procedures in optical see-through head-mounted displays.IEEE transactions on visualization and computer graphics · 2026
- Off-axis microsurgical training using combined exoscopy and endoscopy in a cerebral aneurysm model.Surgical neurology international · 2026
- 3D Reconstruction from Arthroscopic Images using NeRF: a preliminary in-silico studyarXiv · 2026