Early mechanical failure patterns after femoral neck fixation with the femoral neck system and dynamic hip screw.
Clémence Baetz, Adrien Maupetit, Clément Marc, Louis Rony, Guillaume David
Orthopedic Department, Centre Hospitalier Universitaire d'Angers, Angers, France. · Orthopedic Department, Centre Hospitalier Universitaire d'Angers, Angers, France. louis.rony@chu-angers.fr.
Key points
From the paper's conclusion- Within a 24-hour treatment pathway, further reduction of time to surgery was not associated with improved survival in this cohort. Mortality was strongly associated with perioperative medical complications. These findings support current early-surgery targets but suggest that optimisation of perioperative and orthogeriatric care may be at least as important for improving outcomes as further reductions in time to surgery.
Abstract
BACKGROUND: The Femoral Neck System (FNS) has been introduced as a minimally invasive alternative to the Dynamic Hip Screw (DHS) for intracapsular femoral neck fracture fixation. Comparative data on early mechanical failure patterns remain limited. PURPOSE: To describe the early mechanical complications observed after FNS and DHS fixation during a period of progressive FNS adoption, without assuming implant-specific causality. METHODS: This retrospective single-center cohort included 130 consecutive patients treated between December 2020 and January 2024 (42 DHS, 88 FNS). The primary outcome was mechanical revision within one year. Tip-apex distance (TAD) and postoperative reduction quality using the Garden Alignment Index (GAI) were assessed on immediate postoperative radiographs. RESULTS: Mechanical revision occurred in 3 DHS patients (7.1%; 95% CI 1.5-19.5%) and 8 FNS patients (9.1%; 95% CI 4.0-17.1%; p = 0.25). Reduction was acceptable in 40/42 DHS and 80/88 FNS patients (p = 0.50). Mechanical failure occurred in 5/10 patients with unacceptable reduction versus 6/120 with acceptable reduction (p < 0.001). Mean TAD was similar between groups (18.8 ± 4.8 versus 19.6 ± 5.2 mm; p = 0.34), and all four cut-outs occurred with TAD >25 mm. Observed failure patterns included shortening/medialization and cut-out after DHS, and shortening/medialization, cut-out, and peri-implant fracture after FNS. CONCLUSION: Different mechanical complications were observed after DHS and FNS fixation, but confounding by indication and the small number of events preclude implant-specific causal conclusions. Reduction quality and implant positioning appear important determinants of early mechanical failure.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative outcomes of cannulated screws, dynamic hip screws and the femoral neck system for transcervical femoral neck fractures. a retrospective cohort study.Orthopaedics & traumatology, surgery & research : OTSR · 2026
- Characteristics of the Extent and Onset of Osteonecrosis of the Femoral Head After Femoral Neck System (FNS) Fixation: A Minimum Two-Year Follow-Up Comparative Study with Cannulated Screws.Journal of clinical medicine · 2026
- Reduction Quality Determines Failure Behavior in Stable Intertrochanteric Fractures: A Mesh-Free Simulation Comparing DHS and TFNA.JB & JS open access · 2026
- Comparison of internal fixation outcomes in young femoral neck fractures: three cannulated compression screws versus the femoral neck system.Frontiers in surgery · 2026
- Femoral Neck System for young and middle-aged adult femoral neck fractures: a trade-off between reduced lateral irritation and increased neck shortening.Hip international : the journal of clinical and experimental research on hip pathology and therapy · 2026
- Dynamic hip screw versus cannulated cancellous screws fixation for osteosynthesis of femoral neck fractures in adults: a retrospective comparative study.International journal of burns and trauma · 2026