Tissue effects of advanced energy devices: a systematic review and descriptive analysis of vascular sealing and lateral thermal damage in human tissue.
D Marshall, N Kahar, C Challand, L Wyld
University of Sheffield, Western Bank, Sheffield, S10 2TN, UK. Djmarshall1@sheffield.ac.uk. · University of Sheffield, Western Bank, Sheffield, S10 2TN, UK.
Key points
From the paper's conclusion- Lateral thermal spread varied widely (0–7.64 mm) depending on the device and tissue examined as well as the testing methodology and definitions of lateral thermal spread.
Abstract
INTRODUCTION: Advanced energy devices such as bipolar and ultrasonic devices are used ubiquitously in all aspects of surgery. Despite studies demonstrating differences between animal and human tissues, existing human studies have only been conducted in terms of operative outcomes such as blood loss, with a paucity of data on vessel sealing and lateral thermal spread. This study aims to consolidate all existing literature on these characteristics for each device in human tissues to inform clinician decision making. METHODS: A systematic review of the literature was carried out using PRISMA guidelines. Papers which discussed vessel sealing or lateral thermal spread with any methodology were included. Exclusion criteria included animal studies, conference abstracts or papers where full text English language papers were not available. Risk of bias was assessed using ROB2, ROBINS-I v2, CASP case-control and JBI case series checklists as appropriate for the study design. A descriptive analysis was undertaken. RESULTS AND DISCUSSION: Across the twenty-eight included studies, most vessel seals with bipolar and ultrasonic devices exceeded physiological burst pressures. Seal failures occurred unpredictably. Lateral thermal spread varied widely (0-7.64 mm) depending on the device and tissue examined as well as the testing methodology and definitions of lateral thermal spread. Due to substantial variation in study methods, direct statistical comparisons were not possible and a reliable device-to-device comparison cannot currently be made from the available literature. This highlights the need for standardized testing methodology. PROSPERO registration and full protocol availability: CRD420251239034.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Vessel Sealing Systems in Veterinary Oral and Maxillofacial Surgery: Current Evidence and Clinical Applications.Journal of veterinary dentistry · 2026
- Plug-based versus suture-based vascular closure devices in large-bore femoral access for endovascular repair: a systematic review and meta-analysisFrontiers in Cardiovascular Medicine · 2026
- Post-market safety signals of energy-based surgical devices: a cross-specialty disproportionality analysis of the FDA MAUDE database with robotic-platform validation and regulatory triangulation, 2015-2025.Journal of robotic surgery · 2026
- Robotic ultrasound-guided biopsy, ablation, and access: a bibliometric analysis and validation-evidence map : Original Article.Journal of robotic surgery · 2026
- Bio-Based Hemostatic Adhesives, Sealants, and Local Hemostatic Agents for Liver Tissue: A Systematic Review with Meta-Analysis of Preclinical Data.Journal of functional biomaterials · 2026
- Comparison of Y20 with GEN11CN (Harmonic HD 1000i) ultrasonic energy device for sealing 5-7 mm pulmonary vessels in thoracoscopic surgery: a multicenter, randomized, assessor- and statistician-blinded, noninferiority trial.Journal of thoracic disease · 2026