Carbon footprint of disposable, hybrid, and reusable trocars in diverse hospital settings: a life cycle assessment.
M M M Eussen, R M Brandwijk, P Nakhate, H Touw, Y van der Meer, T Stobernack, N D Bouvy
Department of Surgery, Maastricht University Medical Center, P. Debyelaan 25, 6229 HX, Maastricht, The Netherlands. myrthe.eussen@mumc.nl. · Aachen-Maastricht Institute for Bio-based Materials (AMIBM), Maastricht University, Maastricht, The Netherlands.
Key points
From the paper's conclusion- Reusable trocars demonstrated the lowest environmental impact in most evaluated settings, outperforming disposable and hybrid alternatives. In certain settings, hybrid trocars also offered environmental advantages. Improving sterilization efficiency and expanding reuse-compatible device design can further reduce the environmental impact of laparoscopic surgery.
Abstract
BACKGROUND: Trocars are essential instruments in laparoscopic surgery, including high-volume procedures such as laparoscopic cholecystectomy. Although reusable trocars are generally assumed to be more environmentally sustainable than disposable or hybrid variants, previous life cycle assessments (LCAs) have relied on manufacturer-supplied data and rarely account for variation in hospital practices. This study provides the first LCA of disposable, hybrid, and reusable trocars based on independently verified material composition and multiple real-world clinical processing settings. METHODS: A cradle-to-grave LCA was performed following ISO 14040/14044 standards. The material composition was determined by Fourier-transform infrared spectroscopy. The functional unit was defined as four trocars (2 × 5 mm, 2 × 12 mm) required for one laparoscopic cholecystectomy. Environmental impacts were evaluated across three hospital settings: an academic hospital with in-house sterilization, an academic hospital with outsourced sterilization, and a general hospital with in-house sterilization. Inventory data were derived from hospital-specific questionnaires, interviews, manufacturer contact, and published sources. Impacts were assessed using ReCiPe 2016 (hierarchical perspective). RESULTS: Reusable trocars consistently showed the lowest carbon footprint across settings (0.33-0.94 kg CO2-eq per procedure), followed by hybrid trocars (0.64-1.50 kg CO2-eq). Disposable trocars had the highest impacts (1.37-1.74 kg CO2-eq). Manufacturing was the dominant source of emissions for disposable devices, whereas sterilization processes dominated for reusable instruments. The contribution of sterilization and waste processing differed between hospital settings. CONCLUSION: Reusable trocars demonstrated the lowest environmental impact in most evaluated settings, outperforming disposable and hybrid alternatives. In certain settings, hybrid trocars also offered environmental advantages. Improving sterilization efficiency and expanding reuse-compatible device design can further reduce the environmental impact of laparoscopic surgery.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Environmental benefits of reintroducing reusable drapes and gowns in the operating room.BJS open · 2026
- Carbon footprint of anaesthesia pharmaceuticals and disposable products during Caesarean delivery: a single-centre prospective observational study.British journal of anaesthesia · 2026
- Balancing trade-offs in circular medical device design: a case study on laparoscopic devices.Journal of industrial ecology · 2026
- Waste production and environmental impact in interventional cardiology: a lifecycle assessment and modelling studyEuropean Heart Journal · 2026
- Environmental Impact of a Vaginal Pelvic Organ Prolapse Procedure: A Life Cycle Assessment with Scenario Modelling of Mitigation Strategies.International urogynecology journal · 2026
- Low-Carbon Medical Device Procurement: A Scoping ReviewSustainability · 2026