Voice prosthesis durability after primary versus secondary tracheoesophageal puncture: a device-level recurrent-event analysis of a retrospective single-centre cohort.
Rafael De Cicco, Rafael Pereira de Souza, Sandra Lopez Calcinez, Victor Bandini Vieira, Tabata Briaunys Milan, Carla Gonçalves Soares Moreira
Department of Head and Neck Surgery, Instituto do Câncer Arnaldo Vieira de Carvalho, São Paulo, Brazil. rafaeldecicco@gmail.com. · Department of Head and Neck Surgery, Instituto do Câncer Arnaldo Vieira de Carvalho, São Paulo, Brazil.
Key points
From the paper's conclusion- The timing of TEP was not associated with voice prosthesis durability. Chemoradiotherapy was associated with shorter device survival, with a wide confidence interval; the study was not powered for equivalence, and confounding by indication cannot be excluded. Device order did not predict subsequent failure. These findings are exploratory.
Abstract
PURPOSE: To compare voice prosthesis durability between primary and secondary tracheoesophageal puncture (TEP), and to examine whether radiotherapy exposure and device order are associated with device survival. METHODS: Retrospective single-centre cohort of 38 consecutive patients (18 primary, 20 secondary TEP) treated between 2006 and 2025. Each prosthesis contributed one interval from insertion to the next replacement; complication-driven replacement was the event, and routine renewals, size adjustments and devices in situ were censored. Radiotherapy exposure was a single mutually exclusive variable (none / radiotherapy alone / chemoradiotherapy). Cox regression used robust standard errors clustered by patient; patient-level rates used negative-binomial regression. RESULTS: In total, 130 device intervals were analysed (57 failures) over 137.0 person-years. Median time to first failure was 21.6 versus 13.6 months (log-rank p = 0.09). Secondary insertion was not associated with failure (adjusted HR 1.05; 95% CI 0.51-2.18). Relative to no radiotherapy, chemoradiotherapy was associated with failure (HR 6.69; 95% CI 1.25-35.83); radiotherapy alone was not. Device order did not predict failure (fourth or later vs. first: HR 0.81; 95% CI 0.43-1.50). Replacement rates were 0.36 and 0.45 per person-year. Leakage caused 45 of 57 failures (78.9%). CONCLUSIONS: The timing of TEP was not associated with voice prosthesis durability. Chemoradiotherapy was associated with shorter device survival, with a wide confidence interval; the study was not powered for equivalence, and confounding by indication cannot be excluded. Device order did not predict subsequent failure. These findings are exploratory.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Factors Associated with Replacement Interval and Complications of Provox Voice Prostheses in Post-Laryngectomy Patients: A Single-Center Retrospective Recurrent-Event Analysis.Cancers · 2026
- Average lifetime of voice prostheses in laryngectomized patients and influencing factors.Acta otorrinolaringologica espanola · 2026
- Advanced Granulation-Related Extrusion and Intratracheal Suspension of a Voice Prosthesis Mimicking Oesophageal Wall Prolapse: Case ReportJournal of Clinical Medicine · 2026
- Social Factors Affecting Tracheoesophageal Prosthesis Abandonment: Restoring Every Voice.Journal of voice : official journal of the Voice Foundation · 2026
- Near‐Obliterative Bulbar Urethral Stricture 12 Years After Artificial Urinary Sphincter Implantation: A Case ReportIJU Case Reports · 2026
- Temporal trends in patients undergoing artificial urinary sphincter implantation: explantation outcomes and early failure patterns in a tertiary reconstructive referral center.World journal of urology · 2026