Multimodal comparison of 19 single-use flexible ureteroscopes: bench testing, operative video assessment, and surgeon ratings.
Engin Denizhan Demirkıran, Massimo Menozzi, Eugenio Ventimiglia, Giovanalberto Pini, Giorgio Bozzini, Davide Perri
Department of Urology, Zonguldak Bülent Ecevit University School of Medicine, Zonguldak, Türkiye, Turkey. eddemirkiran@gmail.com. · Department of Urology, Azienda Socio Sanitaria Territoriale Lariana, Sant'Anna Hospital, Como, Italy.
Key points
From the paper's conclusion- Contemporary single-use fURSs show substantial heterogeneity, and no single device excelled across every domain. Device selection should weigh several performance characteristics against the intended clinical scenario; whether bench performance predicts surgeon-reported usability remains unresolved and warrants larger, adequately powered studies.
Abstract
PURPOSE: Single-use flexible ureteroscopes (fURSs) differ substantially in optical and mechanical properties, yet comparative evidence remains limited. We performed a multimodal in vitro and clinical assessment of 19 commercially available single-use fURSs to quantify performance variability and examine which device characteristics are associated with surgeon-reported usability. METHODS: Each of 19 models was bench-tested for field of view, laser-fiber safety distance, irrigation flow, deflection, image resolution, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and color discrimination. One surgeon used each device in three separate patients (57 procedures total) and completed a standardized questionnaire after every case; three blinded reviewers scored anonymized intraoperative video. As a secondary analysis, device-level bench metrics were related to patient-level outcomes using a permutation test with correction for multiple comparisons. RESULTS: All domains showed marked variability (resolution 0.447-0.826 lp/mm; baseline deflection 219-312°, lowest 134° with accessories; empty-channel irrigation 27.9-39.0 mL/min; field of view 75-110°). In secondary, exploratory analysis, no bench metric showed an association with surgeon-reported or video-based outcomes surviving correction for multiple comparisons (54 tests, minimum q = 0.20). CONCLUSION: Contemporary single-use fURSs show substantial heterogeneity, and no single device excelled across every domain. Device selection should weigh several performance characteristics against the intended clinical scenario; whether bench performance predicts surgeon-reported usability remains unresolved and warrants larger, adequately powered studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Single-use digital ureteroscopes: which characteristics are essential for the next generation? An international survey.The French journal of urology · 2026
- Comparative performance of single-use flexible ureteroscopes with and without direct in-scope suction: a retrospective analysis.Journal of medicine and life · 2026
- One Decade Since the Advent of the Single-Use Flexible Ureteroscope: What Can We Learn from Device Malfunction Reporting Patterns?Journal of endourology · 2026
- Single-use versus Reusable flexible rhinolaryngoscopes: A retrospective cohort study.Brazilian journal of otorhinolaryngology · 2026
- Toward standardized operative reporting in flexible ureteroscopy for stone disease: a survey among urologists.World journal of urology · 2026
- The global flexible and navigable suction ureteral access sheaths (FANS) survey on utility, practices and future needs in flexible ureteroscopy: a EAU Endourology, PEARLS and IAU collaborationWorld Journal of Urology · 2026