Safety and Efficacy of an Irreversible Electroporation System with Real-Time Electrical and Thermal Monitoring for Liver Tumors: A Bicentric Retrospective Study.
Paul Beunon, Lorenzo Pescatori, Baptiste Bonnet, Olivier Sutter, Paul Janvier, Arthur Petit, Timothée Molango, Genti Xhepa,
Department of Anesthesiology, Surgery and Interventional Radiology, Gustave Roussy, 94805, Villejuif, France. paul.beunon@gustaveroussy.fr. · Interventional Radiology Department, Avicenne Hospital, Bobigny, France.
Key points
From the paper's conclusion- This IRE system appears feasible for liver tumors ineligible for thermal ablation, with a favorable preliminary safety profile and encouraging local control. Real-time electrical and thermal monitoring might be an interesting asset to achieve standardization of IRE procedures.
Abstract
PURPOSE: To assess the safety and efficacy of an irreversible electroporation (IRE) system integrating real-time electrical and thermal monitoring for liver tumor ablation. MATERIALS AND METHODS: This bicentric retrospective study included all consecutive patients treated with this device for primary or secondary liver tumors between April 2022 and August 2024. IRE was selected when surgery or thermal ablation was unsuitable. Patient, tumor and procedure characteristics-including voltage, amperage, interprobe distance, and maximal needle tip temperature-were recorded. Procedures were guided by CT, cone-beam CT or ultrasound. The protocol aimed for a "plateau-current" ≥30 A per needle pair while maintaining probe-tip temperature <50 °C near at-risk structures. Complications were classified using the CIRSE classification. Local tumor progression-free survival (LTPFS) was assessed by MRI every 3 months. RESULTS: 27 patients with 31 liver tumors (15 primary, 16 metastatic; mean diameter 17.9 mm) were treated using 3-6 electrodes. No grade ≥3 complications occurred. Median follow-up was 18.4 months; 6- and 12-month LTPFS rates were 87% and 79%. Interprobe distance >20 mm was associated with recurrence (HR=5.64, p=0.017). Mean peak temperature was 53.3 °C. In cirrhotic livers, mean current was higher (38.9 A vs 32.1 A, p=0.004) and peak temperature lower (49.4 °C vs 56.0 °C, p=0.024). CONCLUSION: This IRE system appears feasible for liver tumors ineligible for thermal ablation, with a favorable preliminary safety profile and encouraging local control. Real-time electrical and thermal monitoring might be an interesting asset to achieve standardization of IRE procedures.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Neoadjuvant and adjuvant nivolumab associated with irreversible electroporation in patients with early hepatocellular carcinoma and high risk of recurrence (NIVOLEP trial)Hepatology · 2026
- Irreversible Electroporation for Colorectal Liver Metastases: Long-term Outcomes from a Transatlantic Multicenter Study.Cardiovascular and interventional radiology · 2026
- An integrated implantable device for IRE-based endoluminal focal treatment and therapeutic efficacy assessment.Biomedical engineering online · 2026
- Adaptive radiofrequency ablation using an Octopus multipurpose electrode for small hepatocellular carcinoma adjacent to critical structures: a prospective single-center pilot feasibility study.Ultrasonography (Seoul, Korea) · 2026
- Intraoperative monitoring of malignant liver tumor microwave ablation using contrast-enhanced ultrasound: A preliminary study.Clinical hemorheology and microcirculation · 2026
- Intrarenal pressure during ex-vivo flexible ureteroscopy: a pilot experimental study in cadaveric kidneys.World journal of urology · 2026