Bench Testing of Multiple Valve-in-Valves Within a Large Housing Self-Expanding Pulmonary Valve.
Sebastian Góreczny, Jacek Kusa, Shakeel Qureshi
Department of Cardiac and Vascular Diseases, St. John Paul II Hospital, ul. Prądnicka 80, 31-202, Krakow, Poland. s.goreczny@szpitaljp2.krakow.pl. · Department of Paediatric Cardiology, Faculty of Medical Sciences in Katowice, Medical University of Silesia in Katowice, Katowice, Poland.
Key points
From the paper's conclusion- This study was performed to demonstrate bench-test feasibility of multiple ViV within a large self-expanding pulmonary platform, the VenusP-valve (Venus Medtech).
Abstract
Lifetime management of patients with large dysfunctional right ventricular outflow tract (RVOT) should incorporate a predefined strategy for valve-in-valve (ViV) treatment. The choice of the initial self-expanding platform may determine the feasibility, equipment and interventional techniques required for subsequent ViV procedures. This study was performed to demonstrate bench-test feasibility of multiple ViV within a large self-expanding pulmonary platform, the VenusP-valve (Venus Medtech). Bench testing was performed using a modular, human-grade anatomical right heart simulator (SimuPULS Pulmonary platform, Simulands). Continuous real-time monitoring was performed using biplane angiographic system and four synchronized cameras. Sequential ViV implantation was performed using progressively smaller VenusP-valves (34 mm, 32 mm, 30 mm) deployed within the initial 36 mm VenusP-valve. Subsequently, a 29 mm balloon-expandable valve (Vitae, Venus Medtech) was implanted within the multi-layered self-expanding valves. Fluoroscopic measurements of the self-expanding valve frame demonstrated minimal under expansion (~ 2 mm) of the tubular section. Final measurements confirmed appropriate expansion of the balloon-expandable valve within the complex of four previously implanted self-expanding valves. The external diameters of the tubular section at peak balloon inflation and after balloon deflation demonstrated minimal recoil (< 1 mm). Final external diameters did not exceed the nominal diameter of the initial 36 mm valve. Valve-in-valve implantation within a large-frame self-expanding valve, enables sequential implantation of progressively smaller self-expanding valves whilst still preserving the option for subsequent implantation of a large balloon-expandable valve. These bench findings may support clinical decision making and ViV planning for percutaneous lifetime management of patients with dysfunctional RVOT.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Case Example and Bench Testing of Hydrodynamic Performance and Migration Stability of Various Valve-in-Valve Configurations within a Large-Housing Self-Expanding Pulmonary Valve.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- First-in-Human Valve-in-Valve Implantation of a Self-Expanding Pulmonary Valve Within a Dysfunctional Balloon-Expandable Pulmonary Valve.JACC. Case reports · 2026
- Early Outcomes of Self-Expandable Versus Balloon-Expandable Valves for Managing Dysfunctional Right Ventricular Outflow TractsCirculation Cardiovascular Interventions · 2025
- Balloon- Versus Self-Expanding Transcatheter Valves for Failed Small Surgical Aortic Bioprostheses: 3-Year Results of the LYTEN TrialCirculation Cardiovascular Interventions · 2026
- Impact of self-expanding valve selection on hemodynamics and clinical outcomes in patients with borderline annular dimensions for balloon-expandable valve: insights from the multicenter-TAVI registryEuropean Heart Journal · 2026
- Outcomes of Supra-Annular vs Intra-Annular Self-Expanding Valves in Valve-In-Valve Transcatheter Aortic Valve Replacement.JACC. Cardiovascular interventions · 2026