Neuroradiology

First-in-human use of SmartGUIDETM steerable 0.014'' microguidewire for the treatment of intracranial aneurysms: a prospective single-center case series.

Jeremy Hofmeister, Gianmarco Bernava, Andrea Rosi, Olivier Brina, Philippe Reymond, Hasan Yilmaz, Michel Muster, Guillaume Petitpierre,

Department of Diagnostic & Interventional Neuroradiology, University Hospital of Geneva, Geneva, Switzerland. jeremy.hofmeister@hug.ch. · Department of Diagnostic & Interventional Neuroradiology, University Hospital of Geneva, Geneva, Switzerland.

Case report · n=5

Key points

From the paper's conclusion
  • This first-in-human prospective series supports the feasibility and safety of the SmartGUIDE™ steerable 0.014″ microguidewire for the treatment of intracranial aneurysms. The study reports reliable navigation to the target without device-related malfunction or complication, favorable operator-reported usability, and highlights the value of incorporating patient-specific flow-models for pre-procedural training. Comparative studies are required to establish efficiency advantage over conventional microguidewires.

Abstract

PURPOSE: To report the first-in-human, prospective clinical evaluation of a steerable 0.014″ microguidewire SmartGUIDE™ (Artiria Medical, Switzerland) for endovascular treatment of intracranial aneurysms, assessing technical feasibility, safety, and operator-reported usability. METHODS: Adult patients undergoing elective treatment of unruptured intracranial aneurysms were prospectively enrolled between August 2023 and July 2024 at a single center. All procedures were performed by a senior neurointerventionalist. The primary endpoint was the device technical success, defined as successful SmartGUIDE™ navigation to a predefined anatomical target. Successful microcatheterization over the wire was a clinically meaningful secondary endpoint, together with safety, device-related adverse events, navigation time, and operator-reported usability (5-point Likert scale). Patient-specific Flow Twin® vascular replicas were used for pre-therapeutic training to refine strategy and mitigate risks inherent to first-in-human use. Clinical and imaging follow-up was performed at 1 year. RESULTS: Five patients were included. SmartGUIDE™ successfully reached the predefined target in all procedures (5/5). Microcatheterization was achieved in four cases (4/5). Median SmartGUIDE navigation time was 7 min (IQR 4-10). No wire-related complications or device malfunctions were observed. Operators reported high device usability, with median Likert scores of 4-5/5 across all parameters. At 1-year follow-up, all aneurysms demonstrated complete and stable occlusion without neurological morbidity. CONCLUSION: This first-in-human prospective series supports the feasibility and safety of the SmartGUIDE™ steerable 0.014″ microguidewire for the treatment of intracranial aneurysms. The study reports reliable navigation to the target without device-related malfunction or complication, favorable operator-reported usability, and highlights the value of incorporating patient-specific flow-models for pre-procedural training. Comparative studies are required to establish efficiency advantage over conventional microguidewires.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers