Towards reducing back-exchange in hydrogen-deuterium exchange mass spectrometry: a feasibility study of sub-zero digestion and nano-LC separation.
Eva Audouin, Isabelle Billault, Guillaume van der Rest
Université Paris-Saclay, CNRS, Institut de Chimie Physique, UMR8000, 91405, Orsay Cedex, France. · Université Paris-Saclay, CNRS, Institut de Chimie Physique, UMR8000, 91405, Orsay Cedex, France. guillaume.van-der-rest@universite-paris-saclay.fr.
Key points
From the paper's conclusion- Overall, these results demonstrate the feasibility of sub-zero protein digestion and its integration with sub-zero nano-LC, while highlighting the trade-off between increased deuterium retention and reduced digestion efficiency.
Abstract
In recent years, hydrogen-deuterium exchange mass spectrometry (HDX-MS) has emerged as the method of choice for characterizing increasingly complex protein assemblies. However, deuterium back-exchange remains a major limitation, as it requires rapid analysis to preserve deuterium labeling and ensure data accuracy. Reducing the temperature during the HDX-MS workflow has been recognized as an effective way to enhance deuterium retention. While sub-zero liquid chromatography (LC) has been successfully implemented, protein digestion at sub-zero temperatures has never been investigated. In this study, we evaluated the feasibility of a complete sub-zero workflow that integrates protein digestion and nano-LC, using 30% ethylene glycol (EG) in the mobile phase to prevent freezing. At - 13 °C, our nano-LC system achieved 20.5% increase in the relative retention of deuterium on peptides compared with 0 °C conditions. We also established the feasibility of performing offline digestion in the presence of EG. Moreover, we demonstrated that pepsin remains active at temperatures as low as - 19 °C, although digestion efficiency decreases under these conditions. Specifically, a 5-min digestion at - 19 °C in the presence of 30% EG resulted in an average increase of 4.7% in relative deuteration compared with a 0 °C digestion for the same duration. Overall, these results demonstrate the feasibility of sub-zero protein digestion and its integration with sub-zero nano-LC, while highlighting the trade-off between increased deuterium retention and reduced digestion efficiency. Further studies are needed to determine whether the additional deuterium retention achieved at sub-zero temperatures translates into improved HDX-MS measurements and structural interpretation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Systematic Evaluation of the Impact of Quench Conditions on Deuterium Back-Exchange in HDX-MSJournal of the American Society for Mass Spectrometry · 2026
- msengine: Adaptive Open-Source Signal Extraction for HDX MSbioRxiv · 2026
- LemonCatcher Acidic Pull-Down Enables Selective In-Cell Hydrogen-Deuterium Exchange Mass SpectrometrybioRxiv · 2026
- Multi-Ensemble Reweighting Enables Structural Interpretation of Differential HDX-MSbioRxiv · 2026
- Droplet Microfluidic Hydrogen/Deuterium Exchange for Investigating Protein Dynamics with Millisecond Precision.Analytical chemistry · 2026
- A Practical Workflow for Reliable MALDI-Based Hydrogen/Deuterium Exchange Mass Spectrometry.Mass spectrometry (Tokyo, Japan) · 2026